Keison KANE200 Manual Download Page 16

14 KANE200 

series 

manual 

 

 

9.  METER SPECIFICATION  

(NOTE MAY BE SUBJECT TO CHANGE)

 

 

Parameter 

 

 

Resolution 

 

Accuracy 

 

Range 

Temp Measurement 

 

Flue Temperature 
 
 
Inlet Temperature 
 
 
Temp (Nett)

 *2

 

 

 
1.0

°

C/F 

 
 
1

°

C/F 

 
 
1.0

°

C/F 

 

 
+2.0

°

C (5

°

F)  

+0.3% reading 
 
+1°C/F  
±0.3% reading 
 
+2

°

C (5

°

F)  

+0.3% reading 

 

 
0-600

°

32-1112

°

 
0-100°C 
32-212°F 
 
0-600

°

C  

32-1112

°

 

 

Gas Measurement

 

Oxygen 
 
*Carbon Monoxide 

(not KANE200) 

 
Carbon Dioxide

*2

 

Efficiency

*2

 

Excess Air

*2

 

 

 

 
0.1% 
 
1ppm 
 
 
0.1% 
0.1% 
0.1% 

 

 

 
+0.2%

*1

 

 
+10ppm <100ppm

*1

 

+5% of reading  
 
+0.3% reading 
+1.0% reading 
+0.2% 

 

 

 
0-21% 
 
0-1000ppm 
 
 
0-30% 
0-99.9% 
0-250% 

 

 

Pre-programmed Fuels 

 

 

KANE250: Natural gas, Light Oil, Propane, Butane, 
LPG. 
KANE200: as above plus heavy oil, coal, anthracite 

 

Dimensions 

Weight 
Handset 
Probe

 

 

 
1kg / 2.2lb 
200mm / 7.9” x 45mm / 1.8” x 90mm / 3.5” 
L300mm / 11.8” x Dia 6mm / 0.25”with 200mm / 7.8” 
long stainless steel shaft, type K thermocouple and 3m 
/ 6ft long neoprene hose 

 

 

Ambient Operating Range 

 

+0

°

C to +40

°

C / 32-104

°

10% to 90% RH non- condensing 

 

Storage Capacity 

 

16 sets of test results in volatile memory 

 

Battery Life 

 

4 AA cells 
>8 hours using Alkaline AA cells 

 

 

AC battery charger 
(optional)  

 

Input:     110Vac/220 Vac nominal 
Output:  10 Vac off load 

 

*1

 

Using dry gases at STP

   

 

*2

 Calculated 

Summary of Contents for KANE200

Page 1: ...KANE200 Efficiency Meter KANE250 Combustion Meter Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG...

Page 2: ...G THE METER AND ITS FOUR BUTTONS 3 4 4 USING THE ROTARY DIAL STARTING FROM MENU 5 6 5 MEASURING FLUE GASES 7 8 6 WHEN YOU FINISH USING THE METER 9 7 METER PROBLEM SOLVING 9 8 METER ANNUAL RECALIBRATIO...

Page 3: ...s The large display shows 2 readings at a time and all data can be printed via an optional infrared printer The printed data can be live data frozen data or stored data 16 sets of tests can be stored...

Page 4: ...e Nickel Metal Hydride NMiH No other battery types are recommended WARNING The battery charger unit must only be used when NMiH batteries are fitted Alkaline batteries are not re chargeable Attempting...

Page 5: ...the flue probe into the bottom of the meter s water trap and connect the probe s temperature plug to the socket next to the water trap check the plug s orientation is correct otherwise incorrect temp...

Page 6: ...meter s ambient temperature will be used as the inlet temperature When countdown is finished the display s top line shows the last selected function and the bottom line displays whatever the rotary di...

Page 7: ...eter normally operates with the pump on Press PUMP quickly to switch the pump off and on When the pump is switched off the meter displays PUMP OFF approx every 30 seconds NOTE the pump will not switch...

Page 8: ...sults when ENTER is pressed DEL ALL clears all the memory when YES and ENTER pressed NOTE ENTER must be pressed to exit this mode VIEW the test number appears on the top line The dial position display...

Page 9: ...te the dial back to MENU and press ENTER before you can continue In STORE View mode the rotary dial is used to select the saved parameter to be displayed Use the UP DOWN keys to change the Test number...

Page 10: ...ircuit it displays EFF Displays calculated efficiency when O2 values are less than 18 EFFn EFFg or EFFc as selected by the user X AIR Displays calculated excess air when O2 values are less than 18 Los...

Page 11: ...be s depth stop cone to fix it in flue diameters from 1 4 to 4 5 inch 6 to 21 mm Its maximum operating temperature is 1112 F 600 C Make sure you do not exceed the meter s operating specifications In p...

Page 12: ...0 series manual EXAMPLE PRINTOUT IN AMBIENT AIR The standard printouts are KANE200 16 char header 16 char header Date 26 12 02 Time 14 27 08 Fuel NATU GAS O2 20 9 CO2 0 0 NETT C AMB C FLUE C EFF G LOS...

Page 13: ...l also include the TEST NO below the header EFF G gross EFF N nett EFF C condensing KANE250 16 char Header 16 char header Date 12 26 02 Time 14 27 08 Fuel NATU GAS O2 20 9 CO2 0 0 CO ppm 00 CO a ppm 0...

Page 14: ...s replacing Oxygen reading CO reading Meter was stored in a cold environment and is not at normal working temperature Oxygen cell or CO sensor needs replacing Pump is switched off Batteries not holdin...

Page 15: ...AND SERVICE Although sensor life is typically more than two years the meter should be re calibrated and serviced annually to stop any long term sensor or electronics drift or accidental damage Local...

Page 16: ...ing 1 0 reading 0 2 0 21 0 1000ppm 0 30 0 99 9 0 250 Pre programmed Fuels KANE250 Natural gas Light Oil Propane Butane LPG KANE200 as above plus heavy oil coal anthracite Dimensions Weight Handset Pro...

Page 17: ...ble of causing interference Check all readings are as expected A level of disturbance is acceptable If not acceptable adjust the meter s position to minimize interference or switch off if possible the...

Page 18: ...ed in mg m3 COn Carbon Monoxide in ppm normalised to the O2R value CO Carbon Monoxide in mg m3 normalised to the O2R value COa Carbon Monoxide calculation as if there were 0 Oxygen present This is onl...

Page 19: ...Time is recorded when each combustion test is printed Note When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be ne...

Page 20: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: