Keison Digilert 100 Operation Manuals Download Page 17

CAUTION:  Do not inject a pulse when the Digilert 100 is turned off.  Do

not exceed 5 volts.

2.  Plug the cable into the upper jack.
3.  Start with the Digilert 100 turned off and the Mode switch set to mR/hr.

Hold down the - button on the end panel while you turn the On/Off/Audio

switch to On.  (Don't use the Audio setting.)
The display shows CAL, and the Digilert 100  counts down for 15 seconds,

chirping each second.  At the end of the 15 seconds, the Digilert 100

beeps several times.

4.  The Digilert 100 collects data for 60 seconds, beeping as it does so, with

CAL and the hourglass indicator flashing.  At the end of the 60 seconds, it

beeps several times. The display shows CAL and SET.  

5.  Use the following table to check the Digilert 100's accuracy. The table

shows appropriate pulse generator count rates to calibrate for Cs137.  If

the accuracy is not within desired limits, follow steps 5-7.  Note that the

Digilert 100 automatically compensates for lost counts due to GM tube

dead time.  Thus, the display reading in CPM mode does not equal the

input frequency.  You can display uncompensated counts in CPM mode by

continuously holding down the - (minus) button; the reading now

corresponds to the input frequency. 

Pulse Generator 

CPM

mR/Hr

μ

Sv/hr           CPS

Input (PPM)

11,629

11,994

10

100

205

22,661

23,888

20

200

399

52,600

59,720

50

500

995

78,541

95,552

80

800

1,593

93,993

119,440

100

1,000

1,991

6. Press the + and - buttons to adjust the reading to what it should be.  When

the reading is correct, press the Set button. The new calibration factor is

automatically calculated from the adjustment you make.

7. The new calibration factor is displayed for several seconds, then the

Digilert 100 beeps and resumes regular operation.

14

Summary of Contents for Digilert 100

Page 1: ...3 06 Nuclear Radiation Monitor Operation Manual...

Page 2: ...Response Times 6 Taking a Timed Total Count 7 Using the Alert 8 The Utility Menu 9 Interfacing to an External Device 10 4 Common Procedures 10 Establishing the Background Count 11 Environmental Area M...

Page 3: ...eriods of time Do not get the Digilert 100 wet Water can damage the circuitry and the coating of the mica surface of the Geiger tube Avoid making measurements with the detector window in direct sunlig...

Page 4: ...ommon applications How the Digilert 100 Detects Radiation The Digilert 100 uses a Geiger Mueller tube to detect radiation The Geiger tube generates a pulse of electrical current each time radiation pa...

Page 5: ...ng the current function and the battery condition The numeric display A shows the current radiation level in the unit specified by the mode switch setting A small battery B appears to the left of the...

Page 6: ...g instead of the current radiation level The current unit of measurement I CPM CPS mR hr or Sv hr is displayed to the right of the numeric display MENU J appears when you are in the Utility Menu The S...

Page 7: ...to set the alert select items in the Utility Menu and set the Calibration settings See Using the Alert and Utility Menu in Chapter 3 and Calibration in Chapter 5 and Buttons Figure 2 5 The and buttons...

Page 8: ...and for users of SI units microsieverts per hour and counts per second To switch between conventional and SI units use the Utility Menu See The Utility Menu in this chapter Starting the Digilert 100...

Page 9: ...0 s reading in mR hr may not be accurate The most immediate indicators of the radiation level are the count light the audio beep and the alert It takes three seconds before an increase is shown on the...

Page 10: ...or 1 mR hr 3 seconds Note You can set the response time to 3 seconds at all radiation levels using the Utility Menu see The Utility Menu in this chapter Taking a Timed Total Count The Digilert 100 ca...

Page 11: ...tes remain 4 At the end of the timed period the Digilert 100 beeps three times and repeats the beeping several times The number displayed is the total count 5 To find the average dose rate for the tim...

Page 12: ...how the MENU icon Release the button and a 1 indicating option 1 will appear along with the Menu icon 2 Scroll through the menu by pushing the or buttons 3 To select an option push the Set button and...

Page 13: ...ide a positive pulse 9 volt peak 1 Kohm impedance each time the Geiger tube detects a count At the tip of the plug the pulse is approximately 1mS wide and is well suited for chart recorders and audio...

Page 14: ...nd look at it from time to time to check for elevated readings You can also use Alert mode to warn you if the radiation increases above the alert level If you suspect an increase in ambient radiation...

Page 15: ...ration The Digilert 100 should be calibrated as often as your regulations require The best way to calibrate is using a calibrated source If no source is available it is possible to calibrate electroni...

Page 16: ...e Digilert 100 collects data for 60 seconds beeping as it does so with CAL and the hourglass indicator flashing At the end of the 60 seconds it beeps several times The display shows CAL and SET You ca...

Page 17: ...ws appropriate pulse generator count rates to calibrate for Cs137 If the accuracy is not within desired limits follow steps 5 7 Note that the Digilert 100 automatically compensates for lost counts due...

Page 18: ...e LCD work the LCD may need to be replaced Display works but no defective Geiger tube look through the window to counts are registered check the mica surface of the tube if it is wrinkled or a break i...

Page 19: ...Digilert 100 requires servicing please contact your distributor or the manufacturer at the following address Do not attempt to repair the Digilert 100 it contains no user serviceable parts and you co...

Page 20: ...f a centimeter They are also called photons The energy of X rays is millions of times greater than that of light and radio waves Because of this high energy level X rays penetrate a variety of materia...

Page 21: ...n passes through the chamber and ionizes the gas it generates a pulse of current The Digilert 100 electronically processes these pulses to display the radiation level Radiation Measurement Units Sever...

Page 22: ...t locations can vary greatly due to soil composition altitude and other factors For example normal background at 10 000 feet might be double that at sea level On an airplane the radiation at 35 000 to...

Page 23: ...mR hr 0 to 50 Beeper sounds the alert CPM 0 to 160 000 Count light Red LED flashes with each count Beeper Operational in Audio mode only can be muted Ports Dual miniature jack sends counts to CMOS com...

Page 24: ...air proof and watertight case while still providing you with access to your Inspector for data regardless of the conditions Gamma detection only Easily use The Inspector through recessed clear plastic...

Page 25: ...and perform statistical analysis on the data received The data is displayed on a graph as well as digital and analog on screen meters and can be saved or printed in various ways including a spreadshee...

Page 26: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: