background image

IceMaster G40.1 – Glycol Chiller/IceBank Instruction Manual  

 

Page 

7

 of 

10

 

www.KegLand.com.au 

Last Updated 29/07/2020 2:58 PM

 

 

Filling the reservoir 

The G40 can be operated with either just water in the reservoir or both water and glycol to reduce 
the freezing point of water. 
 

IMPORTANT: If you are only using water in the reservoir and not glycol then do not set the 
temperature of the reservoir below 2°C. 
 

Fill the reservoir to a total of 20L with water if the reservoir is set at 2°C or an appropriate 
Concentration of glycol solution if dropping the reservoir below 2°C according to the table below. 
Ensure the cooling coils are submerged. The amount of glycol required to be added to the reservoir 
depends on the desired set temperature of the reservoir which is dependent on a number of factors 
including heat ingress into the entire system, insulation of the glycol lines, ambient temperature, the 
number of kegs being cooled and the temperature of the kegs. Use the table below to determine 
what percentage of glycol you require. 
 
Glycol Volume % 

Freezing Point °C 

Specific Gravity at 21°C 

0.0 

1.000 

10 

-3.3 

1.006 

20 

-7.8 

1.015 

30 

-13.3 

1.024 

You can measure the glycol percentage using a calibrated refractometer and comparing to the 
specific gravity at 21°C in the above table. 

We would advise running a 20% glycol solution however, if you do not have a jacketed fermenter or 
have the fermenter in a very warm environment you can run it at a higher glycol concentration. The 
pumps are not designed to be run with a glycol concentration higher than 30% hence it is not 
advised to exceed this glycol concentration. Doing so may damage the pumps.  

 

 

Summary of Contents for IceMaster G40.1

Page 1: ...ycol Chiller IceBank Instruction Manual Page 1 of 10 www KegLand com au Last Updated 29 07 2020 2 58 PM IceMaster G40 1 Glycol Chiller IceBank Instruction Manual KL04114 KegLand Distribution PTY LTD w...

Page 2: ...INTO THE MAINS SOCKET WARNING PERFORM A PRESSURE TEST ON ALL PRODUCT LINES PRIOR TO USE WARNING ENSURE THAT THERE IS ALWAYS ATLEAST 100mm OF CLEARANCE ON EACH SIDE OF THE GLYCOL CHILLER TO ALLOW ADEQU...

Page 3: ...d the IEC cable for the agitator pump and glycol reservoir access flap Run the IEC cable for the agitator pump through the cable outlet and plug it into the external socket as shown to the right Attac...

Page 4: ...mum of 1 hour before turning it on to allow the refrigerant to settle After getting your G40 setup firstly perform a leak test and cooling test on your system to ensure that all connections are water...

Page 5: ...ith a duotight Plug Step 4 Set the pressure on the regulator to 1 5 times the dispensing pressure up to a maximum of 50 psi Step 5 Turn the gas cylinder off and leave to sit for a few hours Then check...

Page 6: ...th the G40 unplugged fill the reservoir with 20L of water ensuring that all cooling coils and the impeller of the agitator pump are submerged Step 2 Attach 10mm ID silicone tube to the Water Out barb...

Page 7: ...ed set temperature of the reservoir which is dependent on a number of factors including heat ingress into the entire system insulation of the glycol lines ambient temperature the number of kegs being...

Page 8: ...erving temperature A single pass through the G40 will be able to drop the temperature of the product by 15 C hence if the storage temperature of the keg is more than 15 C above the desired serving tem...

Page 9: ...of the beverage line between the G40 and the tap Alternatively if the pour is too slow then decrease the length of the beverage line between the G40 and the tap When adjusting your beverage line leng...

Page 10: ...rvoir 6 Attach 10mm ID silicone tube to the Water Out barb and connect this silicone tube to the adjacent Water in barb Run the agitator continuously to keep the reservoir agitated and prevent the res...

Reviews: