1.7 Warnings and cautions
Warnings and Cautions – failure to follow these instructions may pose a risk of injury
Environment
• The product has been designed to function safely when at an
ambient temperature b10ºC and +35ºC.
• Do not use in the presence of flammable gases / liquids, or in an
oxygen rich environment.
• The TonoCare cannot be used in the vicinity of sources known
to cause electromagnetic disturbance (magnetic resonance
imaging, computed tomography, radio-frequency identification,
metal detectors, electronic article surveillance and other
electromagnetic security systems).
• Do not bring the TonoCare into a magnetic resonance environment.
• Keep the front window and nozzle area away from large amounts
of dust or fine particles.
• This product should be used in a room with subdued lighting.
Installation and setup
• Please add mounting hole labels to covering the key holes unless
wall mounting the TonoCare.
• Only mount on wall according to Keeler Instructions.
• Keep out of reach of children.
• Do not mount anything on the docking station other than the
hand unit and forehead rest.
• The mains plug is the means of isolating the device from mains
supply - ensure the mains plug is accessible at all times.
• Do not position the equipment so that is difficult to remove the
mains plug from the wall socket.
• Do not fit mains power adapter into a damaged mains outlet socket.
• Route power cords safely to eliminate risk of tripping to user, or
damage to device.
• TonoCare is not intended to be used with wireless technology.
Do not plug a wireless dongle into the USB port on the docking
station.
Operation and use
• USA Federal law restricts this device to sale by, or order of a
physician.
• The accuracy of IOP measurements is known to be affected by
variations and changes in corneal rigidity due to differences in
corneal thickness, intrinsic structural factors or corneal refractive
surgery. It is recommended that these factors are considered
during IOP measurement.
• Do not use if the product is visibly damaged, and periodically
inspect for signs of damage.
• This product should not be immersed in fluids.
Summary of Contents for Tonocare