9. Specifications and electrical ratings

Guidance and manufacturer’s declaration – electromagnetic immunity

The Jazz Otoscope / Ophthalmoscope is intended for use in the electromagnetic environment specified below. The customer or the user of the 
Jazz Otoscope / Ophthalmoscope should assure that it is used in such an environment.

Immunity test

IEC 60601 Test level

Compliance level

Electromagnetic environment - guidance

Electrostatic 
discharge (ESD)
IEC61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors 
are covered with synthetic material, the relative humidity 
should be at least 30%.

Electrical fast 
transient / burst
IEC61000-4-4

± 2 kV for power supply lines
± 1 kV for input / output lines

Not Applicable

Mains power quality should be that of a typical 
commercial or hospital environment.

Surge
IEC61000-4-5

± 1 kV line(s) to line(s)
± 2 kV line(s) to earth

Not Applicable

Mains power quality should be that of a typical 
commercial or hospital environment.

Voltage dips, short 
interruptions and 
voltage variations on 
power supply
input lines
IEC61000-4-11

<5% 

U

T

 (>95 % dip in 

U

T

) for 0.5 cycle

40% 

U

T

 (60 % dip in 

U

T

) for 5 cycles

70 % 

U

T

 (30 % dip in 

U

T

) for 25 cycles

<5% 

U

T

 (>95 % dip in 

U

T

) for 5 s

Not Applicable

Mains power quality should be that of a typical commercial 
or hospital environment. 

Power frequency 
(50/60 Hz) magnetic 
field
IEC61000-4-8

3 A/m

3A/m

If incorrect operation occurs, it may be necessary to 
position the Jazz Otoscope / Ophthalmoscope further 
from sources of power frequency magnetic fields or 
to install magnetic shielding. The power frequency 
magnetic field should be measured in the intended 
installation location to assure that it is sufficiently low.

Note 

U

T

 is the a. c. mains voltage prior to application of the test level.

Summary of Contents for JAZZ SERIES

Page 1: ...Home Back Next Jazz Instruments Instructions for use Jazz Instruments...

Page 2: ...lacement of LED 6 5 Spare LEDs 6 6 Spare rings p8 7 Ophthalmoscope and accessories 7 1 Purpose 7 2 Lens wheel and correcting lenses 7 3 Apertures and filters p9 7 4 Bulb LED replacement 7 5 Spare Bulb...

Page 3: ...Home Back 3 1 Introduction Thank you for purchasing your Keeler Jazz Instruments Manufactured in compliance with Directive 93 42 EEC for medical products Please read and follow these instructions care...

Page 4: ...nd additional information The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93 42 EEC Medical Device Directive Consult instructions for use...

Page 5: ...for signs of damage Do not use in the presence of flammable gases or an oxygen rich environment This product should not be immersed in fluids Batteries must be inserted as per instructions see sectio...

Page 6: ...opropyl alcohol solution 70 IPA by volume Avoid optical surfaces b Ensure that excess solution does not enter the instrument Use caution to ensure cloth is not saturated with solution c Surfaces must...

Page 7: ...ed thus affecting examination For maximum light yield it is recommended to always insert new high quality batteries on replacement Never immerse handles in fluid Ensure that no fluid or condensation p...

Page 8: ...he ear such as forceps turn magnifying lens approx 2 5X enlargement on otoscope head in anti clockwise direction Replace cover lens in reverse direction 6 4 Replacement of LED Remove instrument head f...

Page 9: ...ses The following correcting lenses are available diopters 0 to 20 and 0 to 20 Readings will be displayed on a lit panel Plus values are displayed in black digits and minus values in red digits 7 Opht...

Page 10: ...n the base of the instrument Warning The bulb LED may be hot 7 5 Spare Bulbs LEDs 1011 P 5530 Jazz Ophthalmoscope Bulb 1011 P 5522 Jazz Ophthalmoscope LED Not available in Europe or USA 7 6 Spare ring...

Page 11: ...ent have to be examined for any reason please return it to your supplier or an authorised dealer in your area Addresses can be supplied on request or visit www keeler co uk Jazz instruments and their...

Page 12: ...cle Large Circle Fixation Star Red free filter Cobalt Blue filter see page 9 Diopters 0 to 20 and 0 to 20 see page 9 Complies with Electrical Safety Medical BS EN 60601 1 2006 Electromagnetic compatib...

Page 13: ...amage Exposure to light from this instrument when operated at maximum intensity will exceed the safety guideline after 30 minutes Halogen bulb The light emitted from this instrument is potentially haz...

Page 14: ...put into service in accordance with the instructions guidance provided herein and are used only in the configuration as supplied The Jazz Otoscope and Ophthalmoscope have been tested and should be use...

Page 15: ...RF emissions CISPR 11 Group 1 The Jazz Otoscope Ophthalmoscope uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nea...

Page 16: ...ould be that of a typical commercial or hospital environment Surge IEC61000 4 5 1 kV line s to line s 2 kV line s to earth Not Applicable Mains power quality should be that of a typical commercial or...

Page 17: ...lds strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of...

Page 18: ...tions equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0...

Page 19: ...eler representatives Your Keeler product is guaranteed for 3 years and will be replaced or repaired free of charge subject to the following Any fault due to faulty manufacture The device has been used...

Page 20: ...the Environmental impact of WEEE Waste Electrical Electronic Equipment and minimise the volume of WEEE entering landfills we encourage at Product end of life that this Equipment is recycled and reuse...

Reviews: