6. Installation and commissioning
Preparing the Cryomatic MKII for use
The Cryomatic MKII consists of the following:
• Cryomatic MKII console
• Footswitch
• Mains cord
• High-pressure gas hose
• Exhaust hose
• Adjustable wrench
• Instructions for use
• 2 spare mains fuses
• Cryo probe(s) must be ordered separately
If any of these parts are missing, contact your distributor
immediately.
Installing Exhaust Hose
Connect the exhaust hose provided from the gas exhaust
connection of the console to a scavenging system or suitably
ventilated area (see page 11).
It is the responsibility of the user to ensure the safe disposition
of exhaust gases.
Connecting the Footswitch
Connect the footswitch to the appropriate connection point on
the rear of the console noting the alignment of the orientation
key (see page 11).
The footswitch can be disconnected for storage and to facilitate
cleaning. Disconnection is achieved by pulling the collar of the
footswitch connector.
Installing the High Pressure Gas Hose
Connect high-pressure hose to the inlet connector (7) at the rear
of the Cryomatic MKII system using the adjustable wrench that
has been provided. Ensure that the coupling is tightened
adequately (see page 13).
Connecting/Changing Gas Cylinders
Gas cylinders must be stored upright, and for a minimum of
eight hours at ambient room temperature prior to use.
Ensure that the gas cylinder is secured properly before use.
Use the following procedure for
connecting or changing gas cylinders: