6
Diagnostics
6.1
Errors
Errors
Cause
Measure
No display after pressing the
operating button
Battery is empty
Charge the battery
Beam is not switched off after
a completed measurement in
automatic mode
Leg of the workpiece too
short or
Surface does not allow for
measurement
No measurement possible;
Switch off device by pressing
the operating button
Beam does not leave the
measurement range, so mea-
surement cannot be com-
pleted
Lift stamp higher after bend-
ing process so measurement
range can be exited reliably
Beam meets the sides of the
workpiece in the gap
Make gap wider or shift de-
vice in the upper tool to the
side so that the beam can
come out unobstructed
Beam maps a point instead
of a line
Safety shut-down not re-
sponding
Immediately switch off;
Send in device to service
partner
If an error code (e.g.
Err 1/1 00014 0x00000000
) is shown on the display,
the error can be acknowledged by briefly pressing the operating button. The
message "
No more errors!
" is displayed. By pressing the operating button
again, the device returns to the main menu and can continue to be used. In
case the error appears again, restarting the device is recommended.
Caution
If the error does not go away or cannot be acknowledged, the device must
be sent back to the service partner for diagnostics and repair.
Summary of Contents for KeMes A2 Series
Page 4: ...A2xx Operating instructions V1 01 4 KEBA ...
Page 7: ...A2xx Table of contents Operating instructions V1 01 7 KEBA Index 54 ...
Page 8: ...A2xx Table of contents Operating instructions V1 01 8 KEBA ...
Page 49: ...A2xx EU Declaration of Conformity Operating instructions V1 01 49 KEBA ...
Page 56: ...A2xx Index Operating instructions V1 01 56 KEBA ...