-13-
Initial Set Mode
Turn on the unit white you press the PROGRAM button to set the following programs.
・
Safety Hold Time
・
Key Entry Sound Volume
・
Temperature Calibration
Safety Hold Time
This function is to cease supply of current to muffle heater to prevent negligence of power switch off when a
certain amount of time passed after program finish. Initial value of 5 hours is stored.
・
When the unit power is off, turn the power switch on while you press the PROGRAM button to display
. Press the button to revise the initial value.
・
0~9 hours is available to set. If you store the
the function of Safety Hold Time is off.
・
Press the
OK/PASS
button to proceed to the “Key Entry Sound Volume”.
Key Entry Sound Volume
Press the button to choose key entry with sound or soundless. (The sound volume set of program M and
H chime is unavailable)
Both key entry and error chimes soundless
Key entry chime sounds but error chime soundless
Both key entry and error chimes sound
Initial Set Value
:
・
Press the
OK/PASS
button to proceed to the next item “Temperature Calibration”.
Temperature Calibration
You can change the temperature calibration modules in the range of ± 20 with button.
Initial
Set
Value
:
(+15)
・
The modification of initial set value is usually unnecessary.
Press the
OK/PASS
button to make sure of the initial value set.
Press
the
PROGRAM
button
to
release
the
initial
value
set
mode.