KCI ActiV.A.C. Reference Manual Download Page 2

Finding a Leak Using Seal Check

Finding a Leak

View or Export Therapy History

Changing the Canister

Therapy History provides a chronological log of dates and
times for unit functions. From Clinician Mode Home Screen,
press 

‘Therapy’

then 

‘Next’

then 

‘History’

to access:

1.  Stop V.A.C.

®

Therapy by pressing                      on the 

     Touch Screen User Interface.
     
     

DO NOT turn power Off to the ActiV.A.C.

®

Therapy Unit.

2.  Slide both tubing clamps toward 
the tubing connector. 

See picture 1.

3.  Close both clamps.

4.  Twist and separate tubing connectors.

5.  Depress the Canister Latch Release. 

     

See picture 2

.

6.  Remove the canister from the therapy unit.

     

Dispose of the canister according to institution

     

requirements or State regulations.

7.  Install the new canister on the therapy 

     unit. An audible “click” should be heard 
     when the canister is properly installed. 
     

See picture 3.

8.  If properly installed, the canister cannot be removed by 

     gently pulling the canister directly away from the unit.

9.  Reconnect the dressing tubing.

10.  Open both tubing clamps.

11.  Start V.A.C.

®

Therapy by pressing                    on the Touch 

      Screen User Interface.

     

DO NOT turn power Off to the ActiV.A.C.

®

Therapy Unit.

The 

‘Therapy History’

screen has two options:

 

‘View History’

: View therapy History on screen.

 

‘Export History’

: Access to screens where the Therapy 

History Report can be transferred via USB or IR (Infrared).

Press 

‘View History’

button to access this on-screen therapy

history display:

Holding the           buttons will rapidly scroll through the
report.

When the ActiV.A.C.

®

Therapy Unit detects a significant leak,

the 

Leak Alarm

will activate. (See 

Alerts and Alarms

section of the User Manual for more details on this alarm.)

The 

Seal Check

feature provides an audible tone and bar

graph to assist in finding leaks.

 The audible tone frequency slows down and the bar graph

   decreases in height as the leak is found.

 Orange bar graph indicates a significant leak.

 Green bar graph indicates that the ActiV.A.C.

®

Therapy 

System is operating normally.

Transition point from 
green to orange and 
vice-versa.

Press to turn the Seal 
Audio tone On or Off.

Press

on the Leak Alarm screen to
use the 

Seal Check

Tool to

help find leaks.

Press 

‘Export History’

button on the 

‘Therapy History’

screen to download a Therapy History Report to a USB Flash
drive or transfer via Infrared transfer.

Most leaks occur:

 where the drape meets the skin.

 where the SensaT.R.A.C.

Pad is attached to the drape.

 at tubing connections.

1. Ensure connector between dressing tubing and canister 

   tubing is properly locked.

2. Ensure the canister is securely installed onto the therapy unit.

3. While therapy is on, move your hand and fingers slowly 

   around the edges of the drape and SensaT.R.A.C.

Pad. 

   The bar graph will lower and the frequency of the           
   audible tone (if Seal Audio is On) will decrease when the 
   leak is found.

4. Refer to the Application Instructions provided with 

   V.A.C.

®

Dressings for information on using excess 

   V.A.C.

®

Drape material to seal the leak area.

5. If necessary, contact your local clinician or KCI for 

   assistance or further information.

Alerts

See the Alerts and Alarms section of the ActiV.A.C.

®

User Manual for more details.

Alerts

will be displayed on the Touch Screen User Interface when the

ActiV.A.C.

®

Therapy Unit detects a condition that requires  patient or

caregiver attention. A single audible tone will sound.

Battery Low Alert:

About 2  hours of batter power remains. Recharge

the battery.

Blockage Alert:

Check tubing for closed clamps, kinks, crimps and

blockages. Lower therapy unit and tubing to or below the wound level.

Low Pressure Alert:

Check tubing for closed clamps, kinks, crimps and

blockages. Lower therapy unit and tubing to or below the wound level.

Service Timer Expired Alert:

Contact KCI for instructions.

Press           for more information.

Press           for more information.

Alarms

See the Alerts and Alarms section of the ActiV.A.C.

®

User

Manual for more details.

Alarms

will be displayed on the Touch Screen User Interface

when the ActiV.A.C.

®

Therapy Unit detects a condition that

requires immediate patient or caregiver attention.

Repeating audible tones will sound.

Press                    to silence the audible tone for 2 minutes.

Battery Critical Alarm:

About 30 minutes of battery power

remains. Immediately recharge the battery.

Blockage Alarm Therapy Interrupted:

Check tubing for

closed clamps, kinks, crimps and blockages. Lower therapy unit
and tubing to or below the wound level.

Canister Full Alarm:

Change the canister and restart therapy.

Canister Not Engaged Alarm:

Ensure canister is properly

installed.

Leak Alarm:

Use leak detection procedures and the 

Seal

Check

Tool to help find and repair leak.

Leak Alarm Therapy Interrupted:

A leak has been detected

and not resolved. After fixing leaks, press ‘Reset’ and restart
therapy.

Low Pressure Alarm Therapy Interrupted:

Therapy unit

remains on; however, negative pressure at the wound may be
below set pressure, potentially compromising therapy. Check
tubing for closed clamps, kinks, crimps and blockages. Lower
therapy unit and tubing to or below the wound level.

IMPORTANT NOTE: The ActiV.A.C.

®

Therapy unit will

continue to attempt to deliver therapy as audio and
visual alarms occur. Timely clinician intervention is   
required.

System Error Alarm:

A technical fault has occurred. Turn unit

off and then back on. If the fault persists, contact KCI.

Therapy Inactive Alarm:

Therapy has been off for

15 minutes. Restart therapy.

Pic. 1

Pic. 2

Pic. 3

ActiV.A.C.

® 

Therapy is to be used with SensaT.R.A.C.

disposables

only. Ensure that a cleanly cut 2-3 cm hole (not a slit) is cut
through the drape. Use the ruler provided in the SensaT.R.A.C.

Dressings package to ensure proper hole size. Center the
SensaT.R.A.C.

™ 

Pad opening directly over hole in drape.

Important

Please ensure you correctly position and orient the unit 
in the carry bag to prevent unnecessary canister alarms.
Whilst in the carry bag, the touch screen on the unit
should always be ‘right side up’ and visible through the
‘window’ in the bag.

Please ensure you cut an appropriate size hole in the
V.A.C.

®

drape to accommodate the SensaT.R.A.C.

pad.

This sh ould be a 2.5cm round hole, not a slit or an ‘X’.

If necessary, contact KCI Ireland for assistance or further
information at 

1800

 

980 8880

.