![KCI ActiV.A.C. International User Manual Download Page 44](http://html2.mh-extra.com/html/kci/activ-a-c/activ-a-c_international-user-manual_3768952044.webp)
On/Off
4. Restart therapy by pressing
.
2. If canister is not full, press “Cancel”.
3. If canister is full, change canister and press ‘
’ on this screen to return to the ‘Home
Screen’. See the
section of this manual for additional information.
Reset
Canister Change
To resolve this alarm:
1. Check fluid level of canister by holding the therapy unit so that the Graduated Marks on the
canister are level and parallel to the floor.
This alarm screen appears when the ActiV.A.C. Therapy Unit
detects that the canister is full and should be replaced.
This alarm will be accompanied by a repeating audible tone.
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing
.
®
‘Audio Pause’
Alerts and Alarms (cont.)
Canister Full Therapy Interrupted Alarm
Reset
Canister Full
Therapy Interrupted
WARNING:
If Canister is full, replace and press ‘Reset’.
If not full, press ‘Cancel’. Press ‘?’ for more information
Cancel
Audio Pause
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
Graduated Marks
A full canister is approximately 300 ml.
42
®
A
ctI