Changing the Temperature Scale
Your thermometer is set at the factory to display Fahrenheit (°F)
temperature measurements. If you wish to switch to Celsius (°C) and/or
back from Celsius to Fahrenheit, follow the following steps:
1) Press the Power Button to turn on the thermometer.
2) Open the Battery Door - use your finger to lift the
battery door open.
3) Find inside the battery compartment a small pin
hole with letters C/F nearby.
4) When the display is in the Ready state with <OK>
displayed, insert a paper clip wire into the hole and
press once to switch temperature display.
5) Follow the steps in the previous “How to Take a
Temperature” section of this manual.
Product Specifications
Displayed temperature range:
95.9–107.6 °F (35.5–42 °C)
Measurement time:
1 second
Operating ambient temperature range:
50–104 °F (10–40 °C)
Display resolution:
0.1 °F (also available in °C)
Accuracy:
Non-predictive mode (liquid bath):
98.6 - 102.2 °F
➔
0.2 °F (37 - 39 °C
➔
0.1 °C)
below 98.6 & above 102.2 °F
➔
0.3 °F (below 37 & above 39 °C
➔
0.2 °C)
Predictive mode (clinical) :
0.4 °F (0.2 °C)
Humidity storage:
< 15-95% non-condensing
Battery life:
2 years or 500 measurements
Symbol Explanations:
°C/°F Pinhole
Subject to change without notice.
This appliance conforms to ASTM Standards for Clinical Thermometers:
E1112 and E1965-98 for clinical accuracy, wherever applicable.
Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
IEC 60601-1:2005 Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.
IEC 60601-1-2: General requirements for basic safety and essential
performance –Collateral standard: Electromagnetic compatibility –
Requirements and tests
Note: Do not use this device in the presence of electromagnetic or other
interferenceoutside the normal range specified in IEC 60601-1-2.
LOT
LOT
NUMBER
TYPE BF
EQUIPMENT
DATE OF
MANUFACTURE