background image

Changing the Temperature Scale

Your thermometer is set at the factory to display Fahrenheit (°F) 

temperature measurements. If you wish to switch to Celsius (°C) and/or 

back from Celsius to Fahrenheit, follow the following steps: 
1)  Press the Power Button to turn on the thermometer. 
2)  Open the Battery Door - use your finger to lift the 

battery door open. 

3)  Find inside the battery compartment a small pin 

hole with letters C/F nearby. 

4)  When the display is in the Ready state with <OK> 

displayed, insert a paper clip wire into the hole and 

press once to switch temperature display.

5)  Follow the steps in the previous “How to Take a 

Temperature” section of this manual.

Product Specifications

Displayed temperature range:  

95.9–107.6 °F (35.5–42 °C) 

Measurement time: 

1 second 

Operating ambient temperature range:  

50–104 °F (10–40 °C) 

Display resolution:  

0.1 °F (also available in °C) 

Accuracy:

Non-predictive mode (liquid bath): 

98.6 - 102.2 °F 

 0.2 °F (37 - 39 °C 

 0.1 °C) 

below 98.6 & above 102.2 °F 

 0.3 °F (below 37 & above 39 °C 

 0.2 °C)

  Predictive mode (clinical) : 

0.4 °F (0.2 °C) 

Humidity storage:  

< 15-95% non-condensing 

Battery life:  

2 years or 500 measurements

Symbol Explanations:

°C/°F Pinhole

Subject to change without notice.
This appliance conforms to ASTM Standards for Clinical Thermometers: 

E1112 and E1965-98 for clinical accuracy, wherever applicable.
Portable and mobile RF communications equipment can affect MEDICAL 

ELECTRICAL EQUIPMENT.
IEC 60601-1:2005  Medical electrical equipment – Part 1: General 

requirements for basic safety and essential performance.

IEC 60601-1-2: General requirements for basic safety and essential 

performance –Collateral standard: Electromagnetic compatibility – 

Requirements and tests

Note: Do not use this device in the presence of electromagnetic or other 

interferenceoutside the normal range specified in IEC 60601-1-2.

LOT

LOT 

NUMBER

TYPE BF 

EQUIPMENT

DATE OF 

MANUFACTURE

Summary of Contents for BehindEar V980USSM

Page 1: ...IMPORTANT READ AND SAVE THESE INSTRUCTIONS Model V980USSM V980 USE AND CARE MANUAL BehindEar Gentle Touch Thermometer ...

Page 2: ...emperatures 7 Do not use this thermometer if the operating ambient temperature range is below 50 F 10 C or over 104 F 40 C 8 Do not mount the battery with the positive sign facing down The thermometer will not work 9 This thermometer is intended for household use only 10 Use of this thermometer is not intended as a substitute for consultation with your physician 11 This thermometer is not intended...

Page 3: ... techniques The thermometer calculates heat flow through the skin and by use of a mathematical algorithm estimates the internal body temperature The temperature reading range of the thermometer is from 95 9 F to 107 6 F 35 5 C to 42 C and the reading time is about 1 second Why Measure Behind the Ear The skin area behind the earlobe is over the posterior auricular artery which is a branch of the ca...

Page 4: ...ide where the temperature is taken from Never take temperature from the side on which the patient was laying Use the opposite side been recently exposed to hot or cold air temperatures Wait at least 20 minutes after the patient was moved to a room with normal operating conditions Readings from the right side of the patient s head may differ from readings taken on the left side of the head Therefor...

Page 5: ...surement Make sure that the Start Button is kept pressed until the beep sounds Retake the temperature and ensure the start button is pressed down quickly Be sure to hold the Start Button until you hear the beep Care and Cleaning If the Sensor needs to be cleaned or disinfected push and hold Start Button to move the tip outward from the Washable Ring Gently wipe the Sensor and the Washable Ring wit...

Page 6: ...tion 0 1 F also available in C Accuracy Non predictive mode liquid bath 98 6 102 2 F 0 2 F 37 39 C 0 1 C below 98 6 above 102 2 F 0 3 F below 37 above 39 C 0 2 C Predictive mode clinical 0 4 F 0 2 C Humidity storage 15 95 non condensing Battery life 2 years or 500 measurements Symbol Explanations C F Pinhole Subject to change without notice This appliance conforms to ASTM Standards for Clinical Th...