Kaye LTR -90 Operating Manual
27
Chapter 6. Calibrator Operation
6.1.1b
Cutout
Table 10 shows and describes the menu selections on the Cutout menu.
Figure 12: Cutout Key Sequence
Table 10: Cutout Menu
Item
Function
1
SOFT CUTOUT
and
HARD CUTOUT
are limit parameters that shut off power to the heat
source if the well temperature exceeds the set limit value.
The Soft Cutout is a safety barrier to protect probes from temperatures more than their
specified temperature limits.
The Hard Cutout protects the Calibrator from irreparable internal damage. See “Set Soft
Cutout” on page 31.
Note:
The Soft Cutout is user configurable and should be set within 5°C to 10°C above
the temperature limit of the instrument being calibrated. The Hard Cutout is not
user configurable and is set by the factory to approximately 160°C.
If the cutout is activated because of an excessive well temperature, the Calibrator enters a
“Cutout mode” where power to the heat source is shut off to immediately cool the
Calibrator and instrument. When in Cutout mode, the instrument displays
“CUTOUT”
above the Stability Status indicator. The instrument stays in Cutout mode until the
temperature is below the cutout SETPOINT and the cutout is reset. See “Reset an Over-
Temperature Cutout” on page 31.
66.03°C
CUTOUT
SOFT CUTOUT: 150 °C
HARD CUTOUT: 160 °C
TEMP
SETUP
CUTOUT
MENU
1
Summary of Contents for LTR -90
Page 4: ...ii no content intended for this page...
Page 14: ...Chapter 2 Specifications 8 Kaye LTR 90 Operating Manual no content intended for this page...
Page 24: ...Chapter 4 Parts and Controls 18 Kaye LTR 90 Operating Manual no content intended for this page...
Page 30: ...Chapter 5 General Operation 24 Kaye LTR 90 Operating Manual no content intended for this page...
Page 49: ......