3
Assembly
3.1
Assembly of the SURGmatic S201 L
3.1.1
Assembly instructions
Material numbers for spare parts are listed in the separate spare parts sheet.
See also: Spare parts sheet for the SURGmatic S201 L.
Requirement
All parts must be cleaned thoroughly.
The glued joint must be clean; there must be no traces of grease or adhesive
residue.
Note
Once assembled, the contra-angle handpiece has to be sprayed thoroughly with
KAVOspray.
3.1.2
Assembling the drive insert
Tool
Name
Mounting mandrel
(Mat. no. 0.411.0961)
▶ Place the ball bearing ⑥ on the head drive ⑤.
▶ Insert the segment ④ into the head drive ⑤.
▶ Place the ball bearing ② on the head drive ⑤.
▶ Insert the assembly mandrel into the drive insert from above, insert the slide
③ in the assembled head drive ⑤ and push the new wire ring ① over the as‐
sembly mandrel until the wire ring ① slots into place in the groove.
3.1.3
Assembling the head casing
Tool
Name
Wrench (Mat. no. 0.411.2371)
Universal bit holder
(Mat. no. 1.002.4577)
Technician's instructions SURGmatic S201 L - 1.009.0470
3 Assembly | 3.1 Assembly of the SURGmatic S201 L
9/19