Operating Instruction STERIclave B 2200 / 2200 P
7 Maintenance
|
7.4 Regular total cleaning
7.4 Regular total cleaning
Proper cleaning and care is essential to the error-free operation of the STERIclave
B. The following cleaning and care operations must be carried out on completion of
60 cycles.
Note
"NEED CLEANING" will appear after 60 cycles.
CAUTION
Parts of the device can be hot
Touching hot objects can cause skin burns.
▶ Make sure that the Power button is off. The bottom display shows "OFF".
▶ Only clean the device when it is cold.
7.4.1 Clean the tray holder and trays
Note
Scouring agents must not be used.
▶ Switch on the device.
▶ Wait until the self test is over.
▶ Open the door.
▶ Take the tray holder and trays out of the chamber.
▶ Use a standard detergent to wash the tray holder and trays, rinse thoroughly and
dry.
7.4.2 Clean the drain filter
▶ Remove and clean/replace the drain filter with the filter key
(Mat. no. 1.005.0724).
See also: 3.4 Scope of delivery, Page 19
61/78