4.3.7
Special chair functions
Note
With operating voltages greater than 200 V, the spindle motors can operate simul‐
taneously, i.e., the chair and backrest can move simultaneously in the automatic
program.
With operating voltages under 200 V, the movements in the automatic program
must be sequential due to the power consumption.
A service technician can perform the changeover.
Note
The lift and backrest motors have thermal fuses. The motors shut down at an op‐
erating temperature of 140 °C. The cooling phase lasts approx. 15 minutes. After
the cooling phase is over, the lift and backrest motors are operable. Such temper‐
atures are not reached in normal practice. The shutoff temperature can be reached
when the motors are frequently actuated in presentations and events (approxi‐
mately 8 complete movement cycles).
4.4
Moving dentist's unit
The swinging range of the dentist unit is limited by stops.
Note
Do not pull the dentist unit by the instrument hose.
▶ To adjust the dentist unit height, release the brake, adjust the height, and reset
the brake.
Instructions for use Status 1080 TM
4 Operation | 4.4 Moving dentist's unit
45/74