Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020
6 Preparation methods according to ISO 17664 | 6.5 Drying
14 / 18
Spray the disinfectant on a cloth, then wipe down the medical device and allow the
disinfectant to act according to the instructions of the disinfectant manufacturer.
Note
Follow the instructions for use of the disinfectant.
6.4.2 Manual disinfection of the interior
For the effective re-preparation, the inside of the device must be cleaned automatically
in a cleaning and disinfection unit in accordance with ISO 15883-1.
The interior of this product is not designed for manual disinfection.
6.4.3 Mechanical disinfection of the exterior and interior
Note
Before setting up, screw the SONICflex tips clockwise onto the SONICflex instru‐
ment.
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the cleaner "neodisher®
mediclean", the neutraliser "neodisher® Z" and rinse "neodisher® mielclear" and only
refers to the material compatibility with KaVo products.)
Note
Insert the instrument with tip screwed on into the adapter for interior cleaning in the
thermodisinfector (e.g. Miele accessories for dental silicon adapters).
▶ The program settings and cleansers and disinfectants that must be used can be
found in the instructions for use of the thermodisinfector.
6.5 Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no lon‐
ger visible.
Automatic Drying
The drying procedure is normally part of the disinfection program of the thermodisin‐
fector.
Note
Follow the instructions for use of the thermodesinfector (compressed air quality ac‐
cording to ISO 7494-2).