KaVo SMARTtorque LUX S619 L Instructions For Use Manual Download Page 9

4

   

Start up and shut down

WARNING

Hazard from non-sterile products.

Infection hazard for care provider and patient.

▶ Before first use and after each use, prepare and sterilise the medical device

and accessories accordingly.

WARNING

Disposal of the product in the appropriate manner.

Prior to disposal, the product and accessories must be appropriately prepared or

sterilised if this is necessary.

CAUTION

Damage from soiled and moist cooling air.

Contaminated and moist cooling air can cause malfunctions and lead to prema‐

ture bearing wear.

▶ Make sure that the supply of cooling air is dry, clean and uncontaminated ac‐

cording to ISO 7494-2.

4.1

   

Mount the MULTIflex (LUX) / MULTIflex LED coupling

▶ Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbine hose

and tighten with the wrench.

▶ Rotate the spray ring on the MULTIflex (LUX) / MULTIflex LED coupling in or‐

der to regulate the fraction of water.

4.2

   

Checking the amount of water

CAUTION

Overheating of the tooth due to insufficient amount of cooling water.

Insufficient water can cause the medical device to overheat and damage the

pulp and tooth.

▶ Adjust the water amount for the cooling to a minimum of 50 cm

3

/min!

▶ Check the cooling water channels and clean the water nozzles with nozzle

needle 1.003.8150 according to need.

 

 

 

Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643

SMARTtorque S619 C - 1.008.1642
4 Start up and shut down | 4.1 Mount the MULTIflex (LUX) / MULTIflex LED coupling

7/21

Summary of Contents for SMARTtorque LUX S619 L

Page 1: ...Instructions for use SMARTtorque LUX S619 L 1 008 1641 SMARTtorque Mini LUX S615 L 1 008 1643 SMARTtorque S619 C 1 008 1642 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...the supply hose 11 6 2 Cleaning the spray nozzle 11 7 Reprocessing methods according to ISO 17664 12 7 1 Preparations at the site of use 12 7 2 Cleaning 12 7 2 1 Manual external cleaning 12 7 2 2 Aut...

Page 4: ...e technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the applicable EC directiv...

Page 5: ...can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a sit...

Page 6: ...adle without the cutter or grinder CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced product life The medical device should be cleaned serviced and...

Page 7: ...ant national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed...

Page 8: ...ion and storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very...

Page 9: ...ount the MULTIflex LUX MULTIflex LED coupling Screw the MULTIflex LUX MULTIflex LED coupling onto the turbine hose and tighten with the wrench Rotate the spray ring on the MULTIflex LUX MULTIflex LED...

Page 10: ...and the medical device and check the following pressures Drive pressure 2 8 bar 40 psi Return air 0 5 bar 7 psi Water 0 8 to 2 5 bar 12 to 36 psi 4 4 Checking the O rings CAUTION Missing or damaged O...

Page 11: ...the medical device is securely locked onto the MULTIflex LUX MULTIflex LED coupling by pulling on it Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it backwa...

Page 12: ...ng and remov ing use gloves or a fingerstall to prevent an injury or infection Forcefully press the push button with your thumb and simultaneously insert the cutter or grinder all the way Check that t...

Page 13: ...front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle CAUTION Hazard from insufficient cooling water Overheating of the medical device and damage to the tooth Check th...

Page 14: ...d Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1...

Page 15: ...dance with EN ISO 15883 1 7 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max...

Page 16: ...Vo care system Follow the instructions for use of the disinfectant 7 3 3 Machine disinfection external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are opera...

Page 17: ...er servicing and care Reduced product life Perform regular proper care and maintenance Note KaVo only guarantees that its products will function properly when the care prod ucts used are those listed...

Page 18: ...ROcare PLUS See also Instructions for Use KaVo QUATTROcare PLUS 2124 A Servicing of the clamping chuck KaVo recommends cleaning and servicing the chuck system once a week See also Instructions for Use...

Page 19: ...with EN 13060 ISO 17665 1 CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medical device with KaVo care produ...

Page 20: ...ted from dust with minimum ex posure to germs in a dry dark and cool space Comply with the expiry date of the sterilised items Instructions for use SMARTtorque LUX S619 L 1 008 1641 SMARTtorque Mini L...

Page 21: ...head INTRA 0 411 9911 Cleanpac 10 units 0 411 9691 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 1776 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo...

Page 22: ...ually extend to lamps light conductors made of glass and glass fibres glassware rubber parts and the col ourfastness of plastic parts All liability is excluded if defects or their consequences origina...

Page 23: ...Phone 420 272 654 566 Fax 420 272 654 732 Email info kavo cz Internet www kavo cz Russia KaVo Dental Russland GmbH nab reki Fontanka 130 A 190005 St Petersburg Phone 7 812 331 86 96 Fax 7 812 251 06...

Page 24: ......

Page 25: ...1 008 1927 kb 20111124 03 en...

Reviews: