7 Processing steps in accordance with ISO 17664 | 7.8 Sterilization
7.8 Sterilization
Sterilization in a steam sterilizer (autoclave) in accordance with ISO
17665-1
NOTICE
Improper service and care.
Premature wear and reduced product service life.
▶ Before each sterilization cycle, service the medical device with KaVo care prod-
ucts.
NOTICE
Contact corrosion due to moisture.
Damage to the product.
▶ Remove the product from the steam sterilizer immediately after the sterilization
cycle.
The KaVo medical device has a maximum temperature resistance of up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the following
sterilization processes:
▪ Sterilizer with pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
– Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
– Drying time: 30 min.
▶ Use in accordance with the manufacturer's Instructions for Use.
7.9 Storage
Processed products must be stored, protected from bacteria, to the extent possible,
and dust, in a dry, dark, cool room.
Note
Comply with the expiration date of the sterilized items.