Instructions for use
SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
Page 1: ...Instructions for use SMARTmatic S20 1 011 6750 SMARTmatic S20 K 1 011 6751 SMARTmatic S20 S 1 011 6752...
Page 2: ...Distributed by KaVo Dental Corporation 11729 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...
Page 3: ...g O rings on the motor coupling 16 6 2 2 Cleaning the spray tube 16 7 Processing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Non fixing preliminary cleaning of the...
Page 4: ...8 Tools and consumables 24 9 Terms and conditions of warranty 25 Instructions for use SMARTmatic S20 1 011 6750 SMARTmatic S20 K 1 011 6751 SMARTmatic S20 S 1 011 6752 Table of contents 4 28...
Page 5: ...m Please direct all questions regarding the product service and maintenance to the KaVo Technical Service Toll free 1 888 ASK KAVO 888 275 5286 Email customerservice kavo com Please refer to the seria...
Page 6: ...Air pressure Transportation and storage conditions Humidity limitations Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety notes in this document mu...
Page 7: ...he case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air To test use and remove the tool use a glove or finger guard 2 2 Technical condit...
Page 8: ...g appropriate medical training can injure the patient the user or third parties Make sure that the user has read and comprehends the instructions for use Only employ the device if the user has the app...
Page 9: ...ergonomically shaped The devices can be sterilized by the steam autoclave method Through the tube and the electrical motor connected to a dental unit the SMARTmatic handpieces equipped with a handpie...
Page 10: ...cation described herein You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use In accordance with...
Page 11: ...11 6752 3 Description of the product 3 2 Technical Specifications To use the spray Attach the spray clip to the dental instrument as shown If you use the spray clip attach the hose to the spray clip a...
Page 12: ...the product 3 3 Transportation and storage conditions 12 28 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to 70 C 4 F to 158 F Relative humidity...
Page 13: ...from contaminated and moist cooling air Contaminated and moist cooling air can cause malfunctions Make sure that the supplied cooling air is dry clean and free of contamination in accordance with ISO...
Page 14: ...ing the foot switch while attaching or detaching the medical device Property damage to the medical device Do not connect or remove the medical device while pressing the foot switch Lightly spray O rin...
Page 15: ...ive speed of the tool or abrupt engage ment of the tool Property damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool at a higher speed...
Page 16: ...hile working Service the medical device When the speed drops or is uneven Service the medical device Missing O ring on the motor coupling Replace O ring 6 2 Troubleshooting 6 2 1 Replacing O rings on...
Page 17: ...cessories required Demineralized water 30 C 2 C 86 F 3 6 F Nozzle needle Brush e g medium hard toothbrush Disposable syringe Note Before cleaning take the hose off the spray clip and the contra angle...
Page 18: ...only 7 3 1 Manual external cleaning Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush under flowing tap water 7 3 2 Automated external cleaning KaVo rec...
Page 19: ...nds carrying out an final disinfection of the unpack aged item to guarantee the complete disinfection NOTICE Using the disinfectant bath or chlorine containing disinfectants Functional damage and prop...
Page 20: ...ying Manual drying Clean the outside and inside with compressed air until no drops of water are visi ble Automated drying The drying procedure is usually part of the cleaning program of the washer dis...
Page 21: ...ce with German RKI requirements KaVo recommends servicing the product as part of the processing after each use i e after each cleaning disinfection and before each sterilization Remove the bur Service...
Page 22: ...rvice coupling chuck from the side hatch of the QUATTROcare PLUS and attach it to coupling service station four on the far right A MULTIflex adaptor must be mounted there Press the handpiece together...
Page 23: ...e the product from the steam sterilizer immediately after the sterilization cycle The KaVo medical device has a maximum temperature resistance of up to 138 280 4 F Select a suitable procedure dependin...
Page 24: ...dle 0 410 0931 Spray head INTRA KaVo Spray 0 411 9911 INTRA service coupling 1 009 6143 Spray clip Set 1 012 9230 Silicone hose sterilisable 0 065 5188 Material summary Mat No Adapter INTRAmatic CLEAN...
Page 25: ...not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection inst...