KaVo ProXam Extraoral Installation Quick Manual Download Page 6

1

3

2

2

2

1

2

3

MAINS SUPPLY 

3 x 2,5 mm2 (100-220 VAC, 16A T 250V)

3 x 1,5 mm2 (230-240 VAC, 10A T 250V)

Remove a knockout in the juction box.

Attach strain relief to the junction box

PE

Mains

Junction box

Blue

Brown

Yellow / Green
(Earth)

Blue

Brown

Yellow / Green
(Earth)

X-ray unit power cable

L

L

N
PE

N

CONNECTING MAINS CABLE

L

N

1

2

3

4

KaVo_ProXam_IP5.eps

X-RAY UNITS WITH DETACHABLE

POWER SUPPLY CORD

1

3

2

REMOVE THE TRANSPORT LOCKING BAR

UNSCREW THE TRANSORT LOCKING BAR

REMOVE THE SUPPORT

FRAME AND STYROFOAM

PACKAGING

Summary of Contents for ProXam Extraoral

Page 1: ...installation quick guide KaVo ProXam Extraoral 1 015 1972...

Page 2: ...al tape 9 Metric Allen key set 10 Metric open end key set 11 Torx key TX08 and TX25 1 6 7 8 9 10 11 2 3 4 5 Torx key Torx key TX25 TX08 1000 39 4 Unit dimensions 1300 51 2 Unit dimensions 1730 2390 68...

Page 3: ...3 2 1 2 3 4 5 6 1 2 3 4 5 6 7 1 4 5 3 2 3 A A B B IF B B B Remove Remove 1 3 Adjust Adjust Distance Unscrew 2 Unscrew Tighten Tighten 4 Adjust the adjustable wall support if needed UNPACKING CARDBOAR...

Page 4: ...NOTE The wall support installation must be able to resist a pull out force of 3310 N 338 kg 744 lb 3310 N 338 kg 744 lb Or A B 1 2 2 3 3 1 1 2 3 1 2 3 4 5 6 3 4 5 1 1 IF B 2 B A 3 3 1 2 4 4 ATTACHING...

Page 5: ...wall support installation must be able to resist a pull out force of 3310 N 338 kg 744 lb 3310 N 338 kg 744 lb 1 2 2 3 3 1 125mm 180 3 2 1 1 2 4 Adjust see page2 3 1 1 4 3 5 IF B 2 B A 3 4 2 1 1 ATTAC...

Page 6: ...to the junction box PE Mains Junction box Blue Brown Yellow Green Earth Blue Brown Yellow Green Earth X ray unit power cable L L N PE N CONNECTING MAINS CABLE L N 1 2 3 4 KaVo_ProXam_IP5 eps X RAY UNI...

Page 7: ...NSPORT LOCKING PIN 90kV ROTATING TUBE HEAD EXPOSURE SWITCH INSTALLING STATIONARY EXPOSURE SWITCH CONNECTING EXPOSURE SWITCH CABLE CONNECTING CABLES ON 2D SYSTEMS Romexis PC C a t 6 Clinic network 1 2...

Page 8: ...Reconstruction PC Ethernet switch C a t 6 Cat 6 1 2 3 4 5 6 7 8 Romexis PC Clinic network Cat 6 1 2 3 CONNECTING CABLES ON 3D SYSTEMS ATTACHING COVERS ATTACHING SENSOR HEAD REMOVING ESD TAPE KaVo_ProX...

Page 9: ...n If the X ray unit is equipped with cephalostat Beam check Flat field calibration X ray unit mechanical movements Check the column movements up down Emergency stop button and plate Check the emergenc...