Accessories required:
▪ Tap water 30°C ± 5°C
▪ Brush, e.g. medium-hard toothbrush
▶ Disassemble the gripping sleeve and cannula.
▶ Detach the key part from the gripping sleeve.
▶ Clean the key part, gripping sleeve and cannula by brushing them under run‐
ning water (at least tap water quality).
▶ Immediately remove contamination from materials used in the dental practice
(impression materials, caustic chemicals).
▶ Carefully wipe off contamination from the cannula tip with a soft Q-tip or soft
cloth and alcohol.
CAUTION
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
▶ Use a short nozzle needle (Mat. no. 1.004.4986) to clean the media exit
holes.
▶ Carefully remove contamination from the media exit ports using the short noz‐
zle cleaning needle.
Disinfection of the triple-function and multifunction handpiece
Manual disinfection
External disinfection
▶ Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfection
Not applicable.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
12/35