background image

2

   

Safety

2.1

   

Description of safety instructions

2.1.1

   

Warning symbol

Warning symbol

2.1.2

   

Structure

DANGER

The introduction describes the type and source of the danger.

This section portrays the possible consequences of non-observance.

▶ The optional step covers necessary measures for avoiding hazards.

2.1.3

   

Description of danger levels

Safety instructions with three hazard levels are used in this document for avoiding

personal and property damage.

CAUTION

CAUTION

indicates a hazardous situation that can cause damage to property, or mild or

moderate physical harm.

WARNING

WARNING

indicates a hazardous situation that can cause death or serious injury.

DANGER

DANGER

indicates the maximum hazard level. indicates a directly hazardous situation that

can cause death or serious injury.

2.2

   

Purpose – Proper use

2.2.1

   

General information

The overarching guidelines and/or national laws, national regulations and the rules

of technology applicable to medical devices for start-up and use of the KaVo product

for the intended purpose are to be applied and complied with.

This KaVo product is intended only for use in the field of dentistry. The product may

not be used for a purpose for which it was not intended.

 

 

 

 

 

 

 

 

 

 

 

Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
2 Safety | 2.1 Description of safety instructions

4/15

Summary of Contents for PIEZOlux 0.571.5211

Page 1: ...Instructions for use PIEZOlux 0 571 5211 1 001 8522 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...oad test for early recognition of tip breakage 6 4 Operation 7 4 1 Using the PIEZOlux 7 5 Preparation methods DIN EN ISO 17664 8 5 1 Cleaning 8 5 1 1 Cleaning the instrument 8 5 1 2 Cleaning the grip...

Page 4: ...ons EMC Electro magnetic compatibility 1 1 2 Symbols Refer to chapter Safety Warning Important information for users and technicians KaVo products with this symbol can be sterilised in autoclaves up t...

Page 5: ...m 1 3 Transportation and storage CAUTION It is hazardous to start up the medical device after it has been stored refrigerated This can cause the medical device to malfunction Prior to start up very co...

Page 6: ...ysical harm WARNING WARNING indicates a hazardous situation that can cause death or serious injury DANGER DANGER indicates the maximum hazard level indicates a directly hazardous situation that can ca...

Page 7: ...her brands The PIEZOlux scaler tips are intended for effective and gentle calculus removal The PIEZOlux Paro tips are used in periodontology gentle cleaning PIEZOlux cem tips are used for the insertio...

Page 8: ...ps Insert the tip into the torque wrench and screw it into the handpiece by turning it clockwise The tip is tightly seated when the torque wrench clicks 3 2 Load test for early recognition of tip brea...

Page 9: ...afety precaution 10 N corresponds to weight of 1 kg See also 3 2 Load test for early recognition of tip breakage Page 6 Select the spray on the foot control or the electronics Note The water is suppli...

Page 10: ...unction properly when disinfectants are used that are listed by KaVo since they were tested for proper use on KaVo prod ucts Infection protection For treatments that are a critical treatment according...

Page 11: ...ng CAUTION Ultrasonic cleaning devices Malfunctions from cleaning in ultrasonic cleaning devices Do not place the medical device in an ultrasonic cleaning device Clean the grip sleeve in a thermodesin...

Page 12: ...destruction Make sure that no disinfectants into the device Do not place the medical device in disinfectant baths 5 2 1 Disinfecting the instrument Manual disinfection KaVo recommends the following p...

Page 13: ...ased on material compatibility The mi crobiological efficacy must be ensured by the disinfectant manufacturer Permissible disinfectants AF liquid S M Microcide D rr FD 322 Use disinfectants according...

Page 14: ...ve 140 C 284 F Do not exceed even when drying Immediately remove the parts from the steriliser after the sterilisation cycle 5 3 2 Sterilising the ZEG tips CAUTION Contact corrosion Contact corrosion...

Page 15: ...ition of the contacts is correct Connect the hose coupling to the handpiece and screw it to the hose casing 5 4 2 Care of ZEG tips If needed open the water passage of the tips using a nozzle needle 5...

Page 16: ...ength using the measuring card Mat no 1 001 6958 If the end of the calculus remover tips are in the red range they need to be discarded Dispose of damaged instrument inserts Dispose of worn instrument...

Page 17: ...2 PIEZOlux Perio tip No 63 Mat no 1 000 2222 PIEZOlux Perio tip No 64 Mat no 1 000 2223 PIEZOlux Perio tip No 65 Mat no 1 000 2224 PIEZOlux cem tip No 47 Mat no 0 571 6961 Torque wrench Mat no 1 000 4...

Page 18: ......

Page 19: ......

Page 20: ...1 005 5901 kb 20100125 02 en...

Reviews: