3 Device upgrade
3.6 Remove sensor holder assembly
Remove the sensor holder assembly from the rotating unit. The sensor holder assembly is replaced
with a CEPH imaging compatible mechanism and a CEPH imaging capable sensor.
1.
Disconnect the device cables from i2000 image capture board and the cable clamp and grounding
cable from the sensor holder assembly.
2.
Remove the sensor holder assembly screws (2 pcs on both sides of the rotating unit).
3.
Carefully remove the sensor holder assembly from the rotating unit.
NOTICE!
A new CEPH imaging compatible sensor holder assembly is supplied with the
upgrade kit.
What to do next:
• If the device is
serial number KR1902594
or older, upgrade the linear movement mechanism and
rotating movement motor. See chapters
Linear movement mechanism replacement
on page 19
and
Rotating movement motor replacement
on page 26 for more instructions.
• If the device is installed with PAN/3D wall bracket, continue to disassemble the device. See chapter
Prepare the device for disassembly
on page 27 for instructions.
• If the device is installed with the CEPH wall bracket in an applicable position, skip to chapter
on page 31 and continue the installation.