8 Special procedures
139
https://www.altera.com/download/software/prog-software
The jic/sof programming file for the circuit board to be
programmed, included in the software update package.
8.8.1
Installation procedure
8.8.1.1
R3210, R3300, R3400
1.
Connect the USB-Blaster's USB connector to a PC with
the Quartus II Programmer software installed.
2.
Connect the USB-Blaster's 10-pin connector to the
JTAG port on the circuit board to be programmed (see
circuit board specific connections below).
NOTICE!
See the orientation of the connector in figure
below.
Connect to J1 on the R3210 board (Cephalostat CPU)
Left: Connect to J3304 on the R3300
Right: Connect to J3405 on the R3400
3.
Start Quartus II programmer and click the Hardware
Setup button in the upper left corner.
4.
Select USB-Blaster from the Currently selected hard-
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...