7 Cephalometric calibration and alignment
115
7.2.10
Horizontal adjustment
Rotate the ear holder assembly to access to all locking
screws.
1. Loosen the 4 locking screws (A) on the ear holder as-
sembly.
2. Make adjustments by turning the eccentric cam screw
(C).
Mark the starting position to the scale beside the screw.
The amount of rotation can be seen on the scale.
3. Tighten at least two locking screws when adjustment is
done and verify by taking a test image.
4. When the horizontal adjustment is complete, tighten all
locking screws.
Top view of ear holder assembly.
To ease the loosening, lock the ear holder assembly to its place with the
locking switch.
Locking screw (A)
Adjustment screw (C)
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...