5-5
IEC Command Codes
032-0358-EN Rev B
X-ray On
: Typically equals 1. 0 to (Ticks per frame - 1)
X-ray Off
: Greater than X-ray On and less than (Ticks per frame - 1)
Set as needed for desired length. 50,000 = 1 (ms)
Panel Readout
: 0 to (Ticks per frame - 1). Typically set the same as
X-ray Off, except for continuous X-ray mode.
Pulse Mode
: 0 or 1. If not entered, defaults to 0.
0 = 200 microsecond pulse for panel blanking and valid frame.
1 = either a 5msec pulse for valid frame or 3msec for panel blanking
Tube Range
: If not entered, defaults to zero.
0 = 120 KV tube, 1 = 90 KV tube
X-Ray Commands
XCN
Performs a cine X-ray acquisition for the number of frames requested at the
‘Ticks Per Frame’ rate as setup in the X-ray register.
Command syntax:
XCN
Frames Watchdog
Frames
: 1 to 1024
Watchdog
: 0 = disabled, 1 = enabled
XCT
Perform a CT scan for requested distance, speed, and watchdog settings.
Command syntax:
XCT
Distance Speed Watchdog
Distance
: 1000 to 668563
Speed
: 1000 to 65000
Watchdog
: 0 = disabled (default), 1 = enabled
XIP
Performs a Panoramic scan for the requested parameters.
Command syntax:
XIP
Distance Speed PlatformOffset PlatformStartSpeed
PlatformStopSpeed RampReduction Watchdog
Distance
: 1000 to 668563
Speed
: 1000 to 70000
Platform Offset
: 1 to (Distance - 25000)
Platform Start
: 1000 to 54000
Platform End
: 1000 to 54000
Ramp Reduction
: 0 to 250
Watchdog
: 0 = disabled, 1 = enabled
XP
Performs an exposed X-ray acquisition for the number of frames requested at the
‘Ticks Per Frame’ rate as setup in the X-ray register.
Command syntax:
XP
Frames Watchdog
Frames
: 1 to 1024
Watchdog
: 0 = disabled, 1 = enabled
Miscellaneous Commands
BD
Get Board Information
Command syntax:
BD
Retrieves the board component versioning.
Results returned in format:
OK KRL:x UBT:x FID:x SWB:x
KRL
: Linux kernel version (short)
UBT
: U-Boot version (short)
FID
: 1210 Board Flash (EEPROM) Chip System ID
SWB
: ROMFS Software Build version (HG-ID Date Time)
Summary of Contents for OP 3D Vision
Page 1: ...Service Manual Cone Beam 3D 2D Panoramic Dental Imaging System KaVo OP 3D Vision...
Page 22: ...2 16 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...
Page 48: ...6 8 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...
Page 50: ...7 2 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 1 of 3...
Page 51: ...7 3 System Schematics 032 0358 EN Rev B Schematic Sheet 2 of 3...
Page 52: ...7 4 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 3 of 3...
Page 53: ......