KaVo recommends for example
- STERIclave B 2200/ 2200P by KaVo
- Citomat/ K-series by Getinge
▶
Sterilise without disposable cannula.
Depending on the device, select a suitable method from the following ster‐
ilisation methods.
Autoclave with a triple pre-vacuum for least four minutes at 134°C ± 1°C
(273°F ± 1.8°F)
Autoclave using the gravitational method: sterilise for at least 10 min. at
134°C ± 1°C (273°F ± 1.8°F)
Autoclave using the gravitational for at least 60 min. at 121°C ± 1°C (250°F
± 1.8°F)
Follow the manufacturer's instructions for use.
Preparation methods according to ISO 17664
38
Summary of Contents for Multi-FINEair 333 M
Page 1: ...Instructions for use Multi FINEair 333 M REF 1 001 8003...