Interference immunity
tests
IEC 60601 test levels
Compliance level
Electromagnetic environ‐
ment - Guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV air discharge
± 2/4/6 kV contact dis‐
charge ± 2/4/8 kV air dis‐
charge
Floors should be made of
wood or concrete or be
fitted with ceramic tiles. If
the floor is made of syn‐
thetic material, the rela‐
tive humidity must be at
least 30%.
Fast transient electrical
disturbances/ bursts ac‐
cording to IEC 6100-4-4
± 2 kV for power lines ± 1
kV for input and output
lines
± 2 kV for power lines
The quality of the supply
voltage should corre‐
spond to that of a typical
business or hospital envi‐
ronment.
Surge voltage / surges
according to IEC
61000-4-5
± 1 kV push-pull voltage ±
2 kV common-mode volt‐
age
± 1 kV push-pull voltage ±
2 kV common-mode volt‐
age
The quality of the supply
voltage should corre‐
spond to that of a typical
business or hospital envi‐
ronment.
Voltage interruptions,
short-term interruptions,
and fluctuations of the
supply voltage according
to IEC 61000-4-11
< 5 % V (> 95% interrup‐
tion of the V T for ½ peri‐
od
40% V (60% drop in V T
for 5 periods 70% V (30%
drop in V T for 25 periods
< 5% V (> 95% drop in V
T for 5 s (250 periods)
< 5 % V (> 95% interrup‐
tion of the V T for ½ peri‐
od
40% V (60% drop in V T
for 5 periods 70% V (30%
drop in V T for 25 periods
< 5% V (> 95% drop in V
T for 5 s (250 periods)
The quality of the supply
voltage should corre‐
spond to that of a typical
business or hospital envi‐
ronment. If the user re‐
quires the advanced func‐
tions even in the case of
interruption in the power
supply, we recommend
the feeding of energy to
the from an uninterrupti‐
ble power supply or from
a battery.
Magnetic field at a supply
frequency (50/60 Hz) ac‐
cording to IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
correspond to the typical
values in a business and
hospital environment.
Note:
V T is the alternating mains voltage before the test level is used.
11.3
Guidelines and manufacturer's statement - Electromagnetic
immunity
The operating light is designed for operation in the environment defined below.
The customer or user of the must ensure that it is operated in such an environ‐
ment.
Instructions for use KAVOLUX 1410 C
11 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 11.3 Guidelines
and manufacturer's statement - Electromagnetic immunity
42/44