Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.2 Optical system
▶ Use the mouth mirror Mat. no. 0.236.2507 to check the mirror surfaces of both
the HR mirror and the PR mirror.
▶ If there are any spots or dots on the mirror surfaces, the HR and/or PR mirror
must be removed, cleaned and reinstalled in accordance with the instructions.
See also:
2.2.3 Remove HR/PR mirror, Page 86
2.2.3 Clean the HR/PR mirror surfaces, Page 89
▶ Then perform a complete alignment to assess the energy loss.
Remove HR/PR mirror
Note
Do not exchange the two mirrors with each other.
First remove one mirror, clean it and reinstall it before removing the second mirror.