Technician's instructions KaVoLUX 540 LED U / T
11 Safety checks - Test instructions | 11.2 Ceiling mounting
▪ Type B
▪ The device is firmly connected / threshold: SL < 0,3 Ω
▪ Measurement according to EUL / threshold limit: < 10 mA*
*The EUL threshold is compatible with the value defined in IEC 60601 (DIN EN
60601), taking comment 2 from table 2 into consideration.
Notes on repeat tests
Note
The value determined in these tests must be documented and evaluated together
with the measuring processes. The measured values may not overshoot the spe-
cified values.
Note
Comparisons with previous measurements must be carried out if the measured val-
ues undershoot the threshold values by more than 10 %. The test intervals should
be reduced if a deterioration in values is determined!
11.2.2 Safety check instructions
Visual inspection
Check the following points in advance:
▪ Has the equipment of the ME device or the ME system been changed since the
last inspection?
▪ Was the change documented and approved (test protocol, STK)?
▪ Are there any indications of insufficient safety?
Visual inspection and assessment of the medical device and
accessories
The following list is an example and makes no claim of being complete.
Check the following items:
▪ Stability of the device
▪ No damage to the cladding or casing (cracks, breakage)
▪ Functioning of the carrier systems of the operating light (brakes, height adjust-
ment, etc.)
▪ Condition of the control panels
▪ Condition of the operating light
▪ Connection of the power connection provided by the treatment centre
Check of safety-related labels for legibility and completeness
▶ Check if all safety-related markings (plates and labels) are present and legible.
▶ Check if the rating plate and serial number plates are present and legible.