For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector
(complying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make
sure that the interior and the exterior of the medical device are dry
after completion of the cycle, and then grease it immediately with
servicing agents from the KaVo care system.
7.4 Drying
Manual Drying
▶
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Preparation methods according to ISO 17664
48