Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150
7 Reprocessing steps in accordance with ISO 17664-2 | 7.1 Preparations at the site of use
20 / 28
7 Reprocessing steps in accordance with ISO 17664-2
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
4
Take suitable personal protective measures.
4
Reprocess the medical device right after treatment.
4
The medical device must be dry when transported to reprocessing.
4
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
4
Remove straight and contra-angle handpieces from the medical device.
4
Remove all residual cement, composite or blood immediately.
4
Do not place in solutions or similar substances.
7.2 Manual reprocessing
CAUTION
The dental motor is an integral part of the water-conducting system of
the treatment centre. Germs and bio-films may be formed in the lines.
This is associated with a risk of infection.
This results in the following requirements for the workflow at the practice:
4
Rinse all withdrawal sites of the water-conducting systems for 2 minutes at
the start of the workday (without any transfer handpieces attached).
4
After each treatment of a patient, purge/rinse the cooling water withdrawal
sites with air and/or water for at least 20 seconds.
4
The permanent/intensive germ reduction of the water-conducting systems
is managed through the unit. Please note the compatibility of the materials
with the disinfectants used.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess by hand only.
7.2.1 Manual external cleaning and manual external
disinfection
WARNING
Incomplete disinfection.
Infection hazard.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.