![KaVo INTRA K-LUX 200 Instructions For Use Manual Download Page 17](http://html1.mh-extra.com/html/kavo/intra-k-lux-200/intra-k-lux-200_instructions-for-use-manual_3859458017.webp)
Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting
▶
Insert a new lamp in the recess so that the contact surfaces mate with
those in the holder.
▶
Slide the lamp into the mount. To secure the lamp, turn the ring
③
, mount
the sleeve
②
and engage it in place.
6.2.2 Replacing the O-rings
NOTICE
Improper care of the O-rings.
Malfunction or complete failure.
▶
Do not use Vaseline or other grease or oil.
Note
Stop working if the O-ring is missing or damaged.
Note
The O-rings on the motor attachment may only be lubricated with a cotton
ball which has been wetted with KAVO Spray.
Note
If there is any leakage at the site of coupling to the instrument, always re-
place all 3 O-rings.
▶
Press the O-rings between your fingers to form a loop.
▶
Slide the O-rings forwards and remove them.
▶
Insert new O-rings into the grooves and spray them with KaVo spray.
17 / 26
Summary of Contents for INTRA K-LUX 200
Page 1: ...Instructions for use INTRA K LUX 200 1 000 6717 INTRA K LUX 200 F 1 002 1912 ...
Page 24: ......
Page 25: ......
Page 26: ...1 001 1050 bd 20190408 04 en ...