6.4
Drying
Manual Drying
In general, use dry, clean and uncontaminated cooling air according to EN ISO
7494-2 for manual drying.
▶ Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermodisin‐
fector.
▶ Follow the instructions for use of the thermodisinfector.
6.5
Care products and systems - Servicing
WARNING
Sharp cutters or grinders in the medical device.
Risk of injury from sharp or pointed cutters or grinders.
▶ Remove cutter or grinder.
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Note
KaVo only guarantees that its products will function properly when the care prod‐
ucts used are those listed as accessories, as they were tested for proper use on
our products.
6.5.1
Servicing with KaVo Spray
KaVo recommends servicing the project twice daily (at noon and in the evening
after hours), after each time the machine is cleaned, and before each sterilisation.
▶ Remove the cutter or grinder and close the chuck.
▶ Cover the product with the Cleanpac bag.
▶ Place the product on the cannula, and press the spray button for one second.
Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800
6 Preparation methods according to ISO 17664 | 6.4 Drying