First use

WARNING

Hazard from non-sterile products.

Infection hazard for care provider and patient.

▶ Before first use and after each use, prepare and sterilise the medical device

and accessories accordingly.

CAUTION

Damage from soiled and moist cooling air.

Contaminated and moist cooling air can cause malfunctions and lead to prema‐

ture bearing wear.

▶ Make sure that the supply of cooling air is dry, clean and uncontaminated ac‐

cording to ISO 7494-2.

4.1

   

Attach the MULTIflex coupling

▶ Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbine hose

and tighten with the wrench.

The water fraction of the spray can be controlled through rotating the spray ring

on the MULTIflex (LUX) / MULTIflex LED coupling.

4.2

   

Check the amount of water

CAUTION

Overheating of the tooth due to lack of cooling water.

Thermal damage to the dental pulp.

▶ Adjust the water amount for the spray cooling to a minimum of 50 cm

3

/min

4.3

   

Check the pressure

CAUTION

Compressed air connection on devices.

Dirty and moist compressed air causes premature bearing wear.

▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance

with EN ISO 7494-2.

▶ Insert the test gauge (Mat. no. 0.411.8731) between the

MULTIflex (LUX) / MULTIflex LED coupling and the medical device, and check

the following pressures:

- Drive pressure: 2.8 bar (40 psi)

- Return air: < 0.5 bar (7 psi)

- Water: 0.8 to 2.5 bar (12 to 36 psi)

 

 

 

 

 

Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800
4 First use | 4.1 Attach the MULTIflex coupling

8/18

Summary of Contents for GENTLEsilence LUX 8000B

Page 1: ...Instructions for use GENTLEsilence LUX 8000B 1 000 0800 GENTLEsilence 8000C 1 000 2800 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...664 11 6 1 Preparations at the site of use 11 6 2 Cleaning 11 6 2 1 Manual cleaning external 11 6 2 2 Automated external cleaning 11 6 2 3 Manual cleaning internal 12 6 2 4 Automated internal cleaning...

Page 4: ...r users and service technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the appl...

Page 5: ...us situation that can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal h...

Page 6: ...ication If the soft tissue in the oral cavity is injured the compressed air may enable sep tic substances to enter the tissue Treatment must be discontinued with instruments operated by compressed air...

Page 7: ...roducts and care systems described therein must be used KaVo recommends specifying a service interval at the dental office for a licensed shop to clean serv ice and check the functioning of the medica...

Page 8: ...t national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed th...

Page 9: ...us to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature o...

Page 10: ...spray ring on the MULTIflex LUX MULTIflex LED coupling 4 2 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water am...

Page 11: ...rings CAUTION Missing or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged Number of available O rings 5 Instructions for use GENTLEsilenc...

Page 12: ...ufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles with the nozzle needle Mat no 0 410 0921 Clean the w...

Page 13: ...uired accessories Tap water 30 C 5 C 86 F 10 F Brush such as a medium hard toothbrush Brush off under flowing tap water 6 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accordan...

Page 14: ...infec tors in accordance with EN ISO 15883 1 6 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents wit...

Page 15: ...agents from the KaVo care system Follow the instructions for use of the disinfectant 6 3 3 Machine disinfection external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 1588...

Page 16: ...nd malfunctions from improper servicing and care Reduced product life Perform proper care regularly Note KaVo only guarantees that its products will function properly when the care prod ucts used are...

Page 17: ...and care unit with expansion pressure for effective cleaning and care KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before e...

Page 18: ...ct Immediately remove the product from the steam steriliser after the sterilisa tion cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure de...

Page 19: ...nsert for turbines 0 411 9902 Cleanpac 10 units 0 411 9691 MULTIflex spray head nozzle 0 411 9921 STERIclave bags 0 411 9912 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 17...

Page 20: ...non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in...

Page 21: ......

Page 22: ...1 003 6890 kb 20101109 02 en...

Reviews: