First use
WARNING
Hazard from non-sterile products.
Infection hazard for care provider and patient.
▶ Before first use and after each use, prepare and sterilise the medical device
and accessories accordingly.
CAUTION
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions and lead to prema‐
ture bearing wear.
▶ Make sure that the supply of cooling air is dry, clean and uncontaminated ac‐
cording to ISO 7494-2.
4.1
Attach the MULTIflex coupling
▶ Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbine hose
and tighten with the wrench.
The water fraction of the spray can be controlled through rotating the spray ring
on the MULTIflex (LUX) / MULTIflex LED coupling.
4.2
Check the amount of water
CAUTION
Overheating of the tooth due to lack of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 50 cm
3
/min
4.3
Check the pressure
CAUTION
Compressed air connection on devices.
Dirty and moist compressed air causes premature bearing wear.
▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
▶ Insert the test gauge (Mat. no. 0.411.8731) between the
MULTIflex (LUX) / MULTIflex LED coupling and the medical device, and check
the following pressures:
- Drive pressure: 2.8 bar (40 psi)
- Return air: < 0.5 bar (7 psi)
- Water: 0.8 to 2.5 bar (12 to 36 psi)
Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800