Product description

GENTLEsilence LUX 8000 BS (

Mat. no. 1.000.5600

)

3.1

   

Technical data

Minimum pressure

2.7 bar

Drive pressure

2.7 to 3.2 bar

recommendation

2,8 bar

Air consumption

47 to 53 Nl/min.

Idle speed

330.000 to 380,000 rpm

Recommended operating pressure

2 to 3 N

Can be plugged onto all Sirona quick couplings.

3.2

   

Transportation and storage conditions

CAUTION

It is hazardous to start up the medical device after it has been stored strongly

refrigerated.

This can cause the medical device to malfunction.

▶ Prior to start-up, very cold products must be heated to a temperature of 20°C

to 25°C (68°F to 77°F).

Temperature: -20°C to +70°C (-4°F to +158°F)

Relative humidity: 5% RH to 95% RH absence of condensation

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

Protect from moisture

 

 

 

 

 

 

 

Instructions for use GENTLEsilence 8000BS - 1.000.5600
3 Product description | 3.1 Technical data

7/19

Summary of Contents for GENTLEsilence 8000BS

Page 1: ...Instructions for use GENTLEsilence 8000BS 1 000 5600 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...rings on the coupling on the supply hose 12 6 2 Cleaning the spray nozzle 12 6 3 Change the water filter 12 7 Preparation methods according to ISO 17664 13 7 1 Preparation at the site of use 13 7 2 Cl...

Page 4: ...of 12 months from the date of the invoice subject to the following conditions In case of justified complaints KaVo will honour its warranty with a free replace ment or repair Other claims of any natu...

Page 5: ...rised by KaVo Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery The dealer purchase date type and...

Page 6: ...NING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Purp...

Page 7: ...ould be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse CAUTION Injury or damage due to wear Irregular running noise significant vibration overhea...

Page 8: ...air and service KaVo products Technicians at KaVo branches throughout the world Technicians specially trained by KaVo To ensure proper function the medical device must be set up according to the re pr...

Page 9: ...nd storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold...

Page 10: ...y of at least 50 cm min for spray cooling 4 2 Check the pressure A drive pressure of at least 2 7 to 2 9 bar 39 to 42 psi is required for operating the medical device 2 8 bar 40 psi is ideal The air c...

Page 11: ...to the cou pling WARNING Release of the medical device during treatment A medical device that is not properly locked can release from the Sirona quick coupling during treatment Before each use check...

Page 12: ...nder all the way Check that the cutter or grinder is seated by pulling on it CAUTION Hazard from defective chucking system The tool can fall out and cause injury Pull on the tool to check if the chuck...

Page 13: ...ury Never touch soft tissue with the head or tip since it may be hot and cause a burn After treatment is over remove the cutter or grinder from the contra angle handpiece since injury and infection ma...

Page 14: ...nd exchange if necessary Use the nozzle needle Mat no 0 410 0921 to free the water passage in the spray nozzles 6 3 Change the water filter CAUTION Hazard from insufficient amount of spray water Insuf...

Page 15: ...l device The medical device must be dry when transported for reconditioning Do not place it in a solution or similar 7 2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in t...

Page 16: ...infect in a thermodisinfector or manually 7 3 1 Manually disinfecting the exterior KaVo recommends the following products based on material compatibility Mikrozid AF made by Sch lke Mayr liquid or clo...

Page 17: ...compressed air until water drops are no longer visible Automatic Drying The drying procedure is normally part of the cleaning program of the thermodisin fector Note Follow the instructions for use of...

Page 18: ...5 2 Care with KaVo QUATTROcare Cleaning and care unit with expansion pressure for effective cleaning and care Remove the cutter or grinder Servicing the product Servicing the clamping chuck KaVo reco...

Page 19: ...TION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the sterilisa tion cycle The medical device has a maximum temperature resistance...

Page 20: ...protected from germs as far as possible and dust in a dry dark cool room Note Comply with the expiry date of the sterilised items Instructions for use GENTLEsilence 8000BS 1 000 5600 7 Preparation met...

Page 21: ...003 6884 Cleanpac 10 units 0 411 9691 STERIclave bags 0 411 9912 Nozzle needle 0 410 0921 Water filter with nipple 1 000 4823 SIRONA spray head for KAVOspray 1 005 8365 KAVOspray 2112 A 0 411 9640 SIR...

Page 22: ......

Page 23: ...1 006 2887 kb 20081009 03 en...

Reviews: