Operating Instruction GENTLEray 982 Bleaching
5 Bleaching fibre

 | 

5.3 Exchanging the bleaching fibre

5.3 Exchanging the bleaching fibre

Note

The bleaching fibre must be exchanged when it is damaged.

The following factors indicate that that the lens or bleaching fibre is damaged or

dirty:

▪ Unclear spot diameter (no clear border of the red target beam)

▪ distinct loss of bleaching performance

Check the surface to see if the lens is dirty. To clean it, follow the directions under

the section "Cleaning the lens".

See also: 4.3 Cleaning the lens, Page 18

CAUTION

Contamination of the incoupling surface of the bleaching fibre

Dirty fibre end surfaces from fingerprints or dust can wear the bleaching fibre.

▶ When the bleaching fibre is not connected to the GENTLEray 980, the incoup‐

ling surfaces (device-side plug) of the bleaching fibre must be covered with the

provided dust caps.

Note

To protect the optical components in the device, the SMA socket in the device must

always be closed (with a bleaching fibre or protective cap).

Removing a defective bleaching fibre

▶ Screw the bleaching sleeve ① off the cap ③ and remove.

▶ Put the cap on the lens connector ②. Do not remove!

▶ Unscrew the bleaching fibre ④ from the lens connector ② and thread it through

the cap ③.

22/27

Summary of Contents for GENTLEray 982 Bleaching

Page 1: ...Operating Instruction GENTLEray 982 Bleaching Always be on the safe side...

Page 2: ...Sales KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...3 Setup methods according to DIN EN ISO 17664 10 3 1 Cleaning 10 3 1 1 Cleaning instructions 10 3 1 2 Manual cleaning 11 3 1 3 Machine cleaning 11 3 2 Disinfection 12 3 2 1 Manual disinfection 12 3 2...

Page 4: ...A Instructions for use PA Care instructions MA Assembly instructions TA Instructions for technicians STK Safety check IEC International Electrotechnical Commission RA Repair instructions EMC Electroma...

Page 5: ...982 Bleaching 1 User instructions 1 2 Service 1 2 Service Service hotline 49 7351 56 2800 Service Laser kavo com Please indicate the product serial number in all requests Additional information can b...

Page 6: ...gross negligence or intent this shall only apply in the absence of mandatory legal regulations to the contrary KaVo cannot be held liable for defects and their consequences due to natural wear improp...

Page 7: ...Damage in transit In Germany If external damage to the packaging is visible upon delivery follow the procedure below 1 The recipient must record the loss or damage in the notice of delivery The re cip...

Page 8: ...ign the notice of delivery The recipient can only assert damages against the transportation company ba sed on these records 2 Leave the product and packaging unchanged 3 Do not use the product If the...

Page 9: ...airs The symbols printed on the outside are for transportation and storage and have the following meaning Transport upright with the arrows pointing upwards Fragile protect against knocks Keep dry Max...

Page 10: ...CAUTION CAUTION indicates a hazardous situation that can lead to property damage or minor to mo derate injury WARNING WARNING indicates a hazardous situation that can lead to serious injury or death...

Page 11: ...cations General information Patients who suffer from photodermatosis and photosen sitive patients photoallergies The preferential therapy must be carefully considered in light of the patient s individ...

Page 12: ...ld be carried out directly after each use CAUTION Product damage from improper disinfection Malfunctions Use disinfectant in accordance with manufacturer s instructions Only disinfect by wiping Do not...

Page 13: ...plete surface of the handpiece and bleaching fibre with a clean cotton cloth saturated with isopropanol 70 until no visible contamination re mains Unscrew the bleaching sleeve from the cap and seal th...

Page 14: ...per disinfection Liquids can enter the handpiece Only disinfect the handpiece by wiping Do not disinfect by spraying 3 2 1 Manual disinfection The following disinfectant is permissible Isopropanol 70...

Page 15: ...p Screw the bleaching sleeve off the cap and remove Put the bleaching sleeve in a bag Note The lens connector including the lens and bleaching fibre may not be sterilised They must be cleaned or disin...

Page 16: ...mination of the lens and incoupling and outcoupling surfaces of the bleaching fibre Contamination can damage the bleaching handpiece the bleaching fibre and the GENTLEray 980 diode laser Always check...

Page 17: ...ay 982 Bleaching 3 Setup methods according to DIN EN ISO 17664 3 5 Packaging and storage 3 5 Packaging and storage Store the bleaching sleeve in a bag Always cover the lens and the bleaching fibre wit...

Page 18: ...4 Bleaching handpiece 982 Note The bleaching handpiece 982 Mat no 1 006 2098 may only be used together with the bleaching fibre Mat no 1 005 2896 4 1 Description Bleaching sleeve Lens holder Lens con...

Page 19: ...ndividual parts may only be screwed hand tight Do not use any tools Screw the bleaching sleeve off the cap and remove Push the cap on the bleaching fibre Screw off the lens fastener from the lens hold...

Page 20: ...amage the lens surface and impair the laser beam quality and intensity Make sure that the outside and inside of the lens remains clean Note To protect the bleaching fibre close it with the lens holder...

Page 21: ...ns connector 4 4 Mount the lens connector Carefully place the lens on the lens holder with the flat side down Screw the lens fastener to the lens and the lens holder Make sure that it screws on easily...

Page 22: ...y the silver sleeve CAUTION The bleaching fibre should not be sterilised This can damage the bleaching fibre Clean and disinfect the bleaching fibre when it is becomes dirty CAUTION Contamination of t...

Page 23: ...ibre to the bleaching handpiece 982 5 2 Connecting the bleaching fibre to the bleaching handpiece 982 Push the cap on the bleaching fibre Remove the dust cap Screw the bleaching fibre hand tight to th...

Page 24: ...leaning the lens Page 18 CAUTION Contamination of the incoupling surface of the bleaching fibre Dirty fibre end surfaces from fingerprints or dust can wear the bleaching fibre When the bleaching fibre...

Page 25: ...Instruction GENTLEray 982 Bleaching 5 Bleaching fibre 5 3 Exchanging the bleaching fibre Installing the bleaching fibre See also 5 2 Connecting the bleaching fibre to the bleaching handpiece 982 Page...

Page 26: ...n the pulp Do not perform bleaching at a higher output KaVo accepts no liability for da mage arising from the bleaching procedure Important factors for selecting the bleaching gel The bleaching gel mu...

Page 27: ...section These instructions can vary depending on the bleaching gel The instructions of the bleaching manufacturer are authoritative when in doubt see the safety data sheet Identify the tooth discolor...

Page 28: ...he software version for the GENTLEray 980 the set output can vary Comply with the manufacturer s instructions for using the bleaching gel failing which use a maximum output of 1 6 W The beam diameter...

Page 29: ...edure 6 2 Bleaching Record the initial status and end result with pictures to clarify and confirm the success of the treatment Note These parameters are already preset when selecting the indication Bl...

Page 30: ......

Page 31: ...1 006 4418 kb 20080626 01 en...

Reviews: