Operating Instruction GENTLEray 982 Bleaching
5 Bleaching fibre
|
5.1 Description
5 Bleaching fibre
5.1 Description
Note
The bleaching fibre (Mat. no. 1.005.2896) is only approved and suitable for use
with the GENTLEray 982 bleaching handpiece.
① Handpiece plug - outcoupling surface
② Device plug - incoupling surface
(identifiable by the silver sleeve)
CAUTION
The bleaching fibre should not be sterilised
This can damage the bleaching fibre.
▶ Clean and disinfect the bleaching fibre when it is becomes dirty.
CAUTION
Contamination of the bleaching fibre on the incoupling and outcoupling surfaces
Dirty fibre end surfaces from fingerprints or dust can wear the bleaching fibre.
▶ When the bleaching fibre is not connected to the GENTLEray 980 or bleaching
handpiece 982, always cover the incoupling and outcoupling surfaces with the
provided dust caps.
20/27