CAUTION
Contact corrosion from moisture.
Damage to product
▶ After the sterilization cycle, immediately remove the product from the
steam sterilizer .
The KaVo medical device has a maximum temperature resistance up to
138 ℃ (280.4 °F).
Select a suitable procedure (depending on the available autoclave) from
the following sterilisation processes:
▪ Autoclave with pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
7 Preparation methods according to ISO 17664