![KaVo EXTRAmatic 9040 Operating Instructions Manual Download Page 24](http://html.mh-extra.com/html/kavo/extramatic-9040/extramatic-9040_operating-instructions-manual_658163024.webp)
Operating Instruction EXTRAmatic 9040
4 Start-up
|
4.1 Preparation for commissioning
4 Start-up
4.1 Preparation for commissioning
Note
Mains connection required.
This unit can be:
▪ Installed as a free-standing unit
Integrated within the workbench
Installed beneath the workbench
Note
Note that the total permissible connected load is 3,000 Watt.
If you are operating the unit with a single extraction point, you do not require a
compressed air connection or diverter module.
Seal any extraction vents that you do not require with a plug: Mat. no. 0.657.0492.
If you are operating the unit with more than one extraction point, a compressed air
connection will be required (inlet pressure of at least 1 bar and max. 6 bar).
The operating pressure for the suction path diverter is set at 0.8 bar via a pressure
reducer.
If the pressure is too high, the rubber membranes can become damaged.
2,5
2,0
1,5
1,0
0,5
0bar