2 Safety | 2.3 Accessories and combination with other equipment

2.3 Accessories and combination with other equipment

Use of un-authorized accessories on the device or un-authorized modifications
to the device can lead to injury.

Only use accessories that have been approved for combination with the
product by the manufacturer.

Only use accessories that are equipped with standardized interfaces.

Do not make any modifications to the device unless these have been ap-
proved by the manufacturer of the product.

2.4 Qualification of personnel

Application of the product by users lacking appropriate medical training can in-
jure the patient, the user or third parties.

Make sure that the user has read and comprehends the instructions for use.

Only employ the device if the user has the appropriate medical training.

Comply with national and regional regulations.

Improper use of the device can lead to burns or injuries.

Never touch the handpiece head or handpiece lid to soft tissue!

Do not use the medical device as a light probe.

Use an appropriate light probe for illumination of the oral cavity or site of
preparation.

After treatment, place the medical device properly in the holder, without the
bur.

2.5 Service and repair

Repairs, servicing and safety checks may only be performed by trained service
personnel. The following persons are authorized to do this:

▪ Service technicians of KaVo branches after the appropriate product training

▪ Service technicians of KaVo authorized dealers after the appropriate product

training

Comply with the following items during all servicing work:

Have the service and testing tasks carried out in accordance with the Medi-
cal Product Operator Ordinance.

After servicing, interventions and repairs of the device and before re-use,
have the service personnel perform safety checks on the device.

Following expiration of the warranty, have the tool holding system checked
once a year.

Have the medical device evaluated by a professional shop with regard to its
cleaning, servicing and functional needs according to an in-house service in-
terval. Define the service interval depending on the frequency of use.

7 / 26

Summary of Contents for EXPERTtorque Mini E677

Page 1: ...Instructions for use EXPERTtorque E680 1 006 4700 EXPERTtorque Mini E677 1 006 0100...

Page 2: ...stributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo c...

Page 3: ...amond 13 6 Troubleshooting 15 6 1 Cleaning the spray nozzle 15 6 2 Changing the water filter 15 7 Processing steps in accordance with ISO 17664 16 7 1 Preparations at the site of use 16 7 2 Manual pro...

Page 4: ...vice For scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK...

Page 5: ...conditions Air pressure Transportation and storage conditions Humidity range Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety notes in this docum...

Page 6: ...e appropriately pro cessed and sterilized In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 2 2 Technical condition A damaged...

Page 7: ...the handpiece head or handpiece lid to soft tissue Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of preparation After treat...

Page 8: ...e devices can be sterilized in a steam sterilizer autoclave The EXPERTtorque LUX E680 and the EXPERTtorque Mini LUX E677 are connected to the dental treatment unit by an inside hose and the coupling a...

Page 9: ...otect himself or herself the patient and third parties from danger and avoid contamination from the product 3 2 Technical Specifications NSK PTL CL LED3 Drive pressure bar 2 0 3 0 Drive pressure recom...

Page 10: ...before disposal See also 2 7 Processing steps in accordance with ISO 17664 Page 16 NOTICE Damage from contaminated and moist cooling air Contaminated and moist cooling air can cause malfunctions Make...

Page 11: ...re wear Supply dry clean and uncontaminated compressed air in accordance with ISO 7494 2 only Insert the test manometer Mat no 0 411 8731 between the coupling and the medical device and check the foll...

Page 12: ...e cou pling or reduce its service life Make sure that the high speed air handpiece is accurately coupled and firmly seated on the coupling Accurately attach the medical device to the coupling and push...

Page 13: ...erty damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool at a higher speed than recommended by the manu facturer NOTICE Use of burs wit...

Page 14: ...ERTtorque E680 1 006 4700EXPERTtorque Mini E677 1 006 0100 5 Operation 5 4 Removing the bur or diamond 14 26 After the bur has stopped rotating firmly press the push button with your thumb and simulta...

Page 15: ...it if needed 6 2 Changing the water filter WARNING Hazard from non sterile products Infection hazard for dentist and patient Reprocess and sterilize the medical device properly before the next use CA...

Page 16: ...e the tool Process the medical device as soon as possible after treatment To minimize the risk of infection during processing always wear protective gloves Remove the tool from the medical device Remo...

Page 17: ...the disinfectant manufacturer and proven by an expert opinion CaviCide made by Metrex Consumables required Cloths for wiping the medical device Spray the disinfectant on a cloth then wipe down the me...

Page 18: ...compliance with ISO 15883 1 which are operated using alkaline cleaning agents having a maximum pH value of 10 For program settings as well as cleansers and disinfectants to be used please refer to the...

Page 19: ...m the medical device Position the tip of the spray nipple in the opening and apply the spray Press the spray key once for 1 2 seconds 7 4 2 Servicing with KaVo QUATTROcare PLUS Servicing and cleaning...

Page 20: ...hes back to normal service mode See also 2 Servicing with KaVo QUATTROcare PLUS 7 5 Packaging Note The sterilization bag must be large enough for the handpiece to fit without stretching the bag The qu...

Page 21: ...75 F Drying time 30 min Sterilizer using the gravity method at least 60 minutes at 121 C 250 F Drying time 15 min Remove contra angle handpieces and turbines immediately after the com pletion of the s...

Page 22: ...411 3053 Wrench for lid of EXPERTtorque Mini new 1 008 6133 Wrench for lid of EXPERTtorque Mini old 1 006 3384 Replacement filter 1 002 0271 Wrench 1 002 0321 Instrument stand 2151 0 411 9501 Insert f...

Page 23: ...for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calci...

Page 24: ......

Page 25: ......

Page 26: ...1 009 2488 bd 20171103 03 en US...

Reviews: