KaVo EXPERTmatic E31 C Instructions For Use Manual Download Page 14

Instructions for use EXPERTmatic E31 C – 1.008.4998

5 Operation | 5.4 Removing the dental bur

14 / 28

 WARNING

Use of unauthorised dental burs.

Injury to the patient or damage to the medical device.

4

Comply with the instructions for use and the intended use of the dental bur.

4

Only use dental burs that do not deviate from the specified data.

 CAUTION

Do not use damaged dental burs.

Risk of injury from swallowing a dental bur that falls out.

4

Do not use damaged dental burs.

4

Do not use dental burs with adherent soiling.

 CAUTION

Contaminated, sharp-edged dental bur.

Patients, users or third parties could be infected by contaminated medical
devices.

4

Use gloves or a finger guard when you test, insert and remove the dental
bur.

CAUTION

Dental bur with damaged, worn or deformed shafts.

Material damage to the chuck system, dental bur is difficult or impossible to re-
move from the chuck system.

4

Never use dental burs with damaged, worn or deformed shafts.

4

Block drive with holder.

4

Screw or push the insert onto/into the head.

4

Remove the holder.

5.4 Removing the dental bur

 WARNING

Rotating dental bur.

Cuts, infection and burn injury.

4

Do not touch the dental bur while it is rotating.

4

Never contact soft tissue with the instrument head.

4

Remove the dental bur from the handpiece after treatment to avoid injury
and infection during storage.

4

Block drive with holder.

Summary of Contents for EXPERTmatic E31 C

Page 1: ...Instructions for use EXPERTmatic E31 C 1 008 4998...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...g for malfunctions and troubleshooting 16 6 1 Checking for malfunctions 16 6 2 Troubleshooting 16 6 2 1 Replacing the O rings on the motor coupling 16 7 Reprocessing steps in accordance with ISO 17664...

Page 4: ...Vo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use ar...

Page 5: ...storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original langu...

Page 6: ...ns for use EXPERTmatic E31 C 1 008 4998 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to materi...

Page 7: ...Reprocess the product and accessories appropriately before disposal 4 Use gloves or a finger guard when you test insert and remove the dental bur 2 2 Improper use Because the operation with an electri...

Page 8: ...sories that are equipped with standardised interfaces 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the product 4 Use original KaVo spare parts...

Page 9: ...r third parties This may res ult in aspiration swallowing of parts and possibly even a risk of suffocation 4 Only use spare parts that comply with the specification for repair original manufacturer sp...

Page 10: ...r the described application by a properly trained user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructi...

Page 11: ...speed via the motor control 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm u...

Page 12: ...uct and accessories before disposal Also refer to 7 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 18 Current packaging law Dispose of and recycle the packaging appropriately in ac...

Page 13: ...securely locked onto the coupling CAUTION Removing and attaching the medical device while the drive motor is ro tating Damage to the catch pin 4 Never attach or remove the medical device while the dr...

Page 14: ...ird parties could be infected by contaminated medical devices 4 Use gloves or a finger guard when you test insert and remove the dental bur CAUTION Dental bur with damaged worn or deformed shafts Mate...

Page 15: ...c E31 C 1 008 4998 5 Operation 5 4 Removing the dental bur 15 28 4 Unscrew it or take it off the insert after it has stopped rotating NOTE For the re use of the inserts please refer to the instruction...

Page 16: ...HOUT using KaVo original spare parts Parts could come loose and cause injury Aspiration swallowing of parts and danger of suffocation 4 Only use spare parts that comply with the specification for repa...

Page 17: ...for use EXPERTmatic E31 C 1 008 4998 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 17 28 4 Spray a lint free cloth with KaVo Spray apply it to the new O rings and in sert them in...

Page 18: ...only KaVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opinion...

Page 19: ...internal and external disinfection Manual cleaning Accessories required Tap water drinking water Brush e g medium hard toothbrush 4 Brush off under flowing tap water at least 10 seconds Cleaning with...

Page 20: ...only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only 7 3 1 pre cleaning Accessories required T...

Page 21: ...ng with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation but no later than after 30 minute...

Page 22: ...F Sterilisation parameters Select a suitable process from the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1...

Page 23: ...ry Mat no Holder 1 004 0596 INTRA Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spr...

Page 24: ...liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connectio...

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...1 008 5577 bd 20240417 03 en...

Reviews: