Instructions for use ESTETICA E70 Vision / E80 Vision
9 Eliminating disturbances
9 Eliminating disturbances
Note
In case of malfunctions, consult the separate instructions for the use and
care of the individual instruments (such as the turbine, motor, camera, Sate-
lec Mini LED, etc.).
Malfunction
Cause
Remedy
Nothing works.
Main switch is off.
▶
Turn on main switch.
Nothing works.
The main fuse is
tripped.
▶
Unplug the unit from the mains.
▶
Check and replace, if required, the main ser-
vice fuse.
The main service fuse is situated next to the
master switch.
▶
Use a screwdriver to open the bayonet lock,
and change the microfuse.
(220, 230, 240 V AC: T 6.3 H
Mat. no.
0.223.2783
);
(100, 110, 120, 130 V AC: T 10 H
Mat. no.
1.007.2529
).
▶
The re-close the bayonet closure with the
screwdriver.
No spray in the instru-
ments.
No spray preselected.
▶
Preselect spray.
Open the spray regula-
tion in the instrument
is closed.
▶
Open the spray regulation in the instrument.
The main water valve
in the office is closed.
▶
Open the main water valve in the office.
The compressor is not
turned on.
▶
Turn on the compressor.
Water in the return air
filter.
The O-rings of the
MULTIflex coupling are
damaged.
▶
Replace all the O-rings on the MULTIflex coup-
ling.
The suction hoses do
not have any suction.
The suction machine is
not turned on or is de-
fective.
▶
Turn on the suction machine or eliminate the
defect.
The slide valves in the
conical sections of the
suction hoses are
closed.
▶
Open the slide valve.
The sieves in the se-
lective valves are
plugged.
▶
Exchange the sieve.
The base switch is ac-
tuated.
▶
Stop actuating the base switch.
169 / 178