Instructions for use KaVo CONNECTbase
2 Safety | 2.8 Data backup
10 / 56
The user is responsible for backing up data.
For security reasons, the database and all document files should be backed up
regularly. A data backup schedule should be prepared.
KaVo recommends daily backup.
Backup copies must be stored in a safe place protected from access of third
parties. The detailed procedure should be defined in a data protection concept.
The user is responsible for implementing the current data protection guidelines.
The KaVo CONNECTbase system creates anonymous folders that can be re-
named by the user on a PC.
2.8 Data backup
Hard disk defects can lead to data loss.
4
Generate backup copies regularly.
4
Keep backup copies in a safe location.
The user is responsible for backing up data.
For security reasons, the database and all document files should be backed up
regularly. A data backup schedule should be prepared.
KaVo recommends daily backup.
2.9 Backup copies
Such a copy is also called a backup. It should not be confused with the export
of data. There are various backup strategies.
There are various options:
▪ Full backup (backup of all data on the hard disk)
▪ Incremental backup (backup only of data that has been changed since the
previous backup)
Full data backups are more time-consuming and require significantly more
memory since the copied amounts of data are much larger than in incremental
backups. Combinations of full and incremental backups are possible as well to
suit the security requirements of the office.
A hard disk failure can also occur during a backup.
KaVo recommends to never overwrite the current backup copy.
The computer technician/system administrator of your practice is familiar with
the current technology for backup copies and making backups.
Backup copies must be stored in a safe place protected from access of third
parties. The detailed procedure should be defined in a data protection concept.
2.10 Electromagnetic fields
Electromagnetic fields might interfere with the functions of implanted systems
(such as pacemakers).
4
Ask patients if they have a cardiac pacemaker or other system implanted
and counsel them about the risks before you start the treatment.
High-frequency communications devices may interfere with medical electrical
devices.
4
Do not use cell phones in medical offices, hospitals or laboratories.
4
Remove electronic devices such as e.g. computer storage media, hearing
aids etc. during operation.
Summary of Contents for CONNECTbase
Page 1: ...Instructions for use KaVo CONNECTbase...
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...1 014 7587 Fk 20220302 02 en...