Instructions for use EXPERTmatic LUX E25 L Planmeca – 1.015.0660
7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging
24 / 28
NOTE
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front flap and start the service procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
Also refer to:
7.4.2 Servicing with KaVo QUATTROcare PLUS, Page 23
7.5 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!
4
Seal the medical device separately in a sterile pack.
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / EN ISO 17665-1
CAUTION
Improper service and care.
Risk of injury.
4
Service regularly with suitable agents.
CAUTION
Contact corrosion due to moisture.
Damage to product.
4
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The medical device has a max. temperature resistance of up to 138
o
C (280.4
o
F).
Sterilisation parameters:
Select a suitable process from the following sterilisation processes (depending
on the available steriliser):
▪ Steriliser with triple pre-vacuum:
▪ at least 3 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
▪ Steriliser using the gravity method:
▪ at least 10 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
▪ at least 30 minutes at 121
o
C -1
o
C / +4
o
C (250
o
F -1.6
o
F / +7.4
o
F)
4
Remove the medical device from the steam steriliser immediately after the
sterilisation cycle is completed.
4
Use according to the manufacturer's Instructions for Use.