background image

Instructions for use

EXPERTmatic LUX E25 L Planmeca – 1.015.0660

Summary of Contents for 1.015.0660

Page 1: ...Instructions for use EXPERTmatic LUX E25 L Planmeca 1 015 0660...

Page 2: ...Distributed by Planmeca Oy Asentajankatu 6 A 00880 Helsinki Finland Tel 358 20 7795 500 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...ng for malfunctions and troubleshooting 17 6 1 Checking for malfunctions 17 6 2 Troubleshooting 17 6 2 1 Replacing the O rings on the motor coupling 17 6 2 2 Changing the water filter 18 7 Reprocessin...

Page 4: ...com planmeca KaVo Technical Service If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for...

Page 5: ...Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original language German Hazard levels The warning...

Page 6: ...Instructions for use EXPERTmatic LUX E25 L Planmeca 1 015 0660 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to material damage...

Page 7: ...disposal 4 In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Use gloves or a finger guard when you test insert and remove t...

Page 8: ...d motors 4 Do not use motors for unauthorised purposes 2 4 Accessories and combination with other equipment Use of non authorised accessories or non authorised modifications of the device could lead t...

Page 9: ...se service intervals where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result o...

Page 10: ...statutory regulations Proper use According to these regulations this product may only be used for the described application by a properly trained user You need to comply with the following the applic...

Page 11: ...RA LUX motors and motors with a connector in accordance with DIN EN ISO 3964 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongl...

Page 12: ...nagement recycling companies Comply with the comprehensive return system KaVo has had its packaging licensed for this purpose Please comply with the regional public waste disposal system CAUTION Damag...

Page 13: ...Instructions for use EXPERTmatic LUX E25 L Planmeca 1 015 0660 4 Startup and shut down 4 1 Checking the amount of water 13 28...

Page 14: ...andpieces being at tached and if there is a risk of contamination from reflux or back suction the system may also need to be rinsed for 20 to 30 seconds after each patient 5 1 Attaching the medical de...

Page 15: ...ISO 1797 type 3 are made of steel or hard metal and meet the following criteria Shaft diameter 1 59 to 1 60 mm 0 0626 in to 0 0629 in Overall length max 25 mm Shaft clamping length at least 12 mm Bla...

Page 16: ...tal bur Material damage to dental bur shaft and chuck system reduction of the service life of dental bur and chuck system 4 Do not operate the dental bur at a higher speed than recommended by the manu...

Page 17: ...use of motor 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and cause injury Aspiration swallowing of parts and danger of suffocatio...

Page 18: ...zard to the dentist and patient 4 Prior to initial startup and after each use reprocess the product and ac cessories CAUTION Overheating of the tooth due to insufficient amount of cooling water Insuff...

Page 19: ...ve the dental bur from the medical device 4 Remove all residual cement composite or blood immediately 4 Do not place in solutions or similar substances 7 2 Manual Reprocessing WARNING Sharp dental bur...

Page 20: ...ining products Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only KaVo recommends the following products based on the compatibility of the ma terials The microbiologic...

Page 21: ...al drying Use KaVo DRYspray for subsequent drying of the air water and gear unit ducts 4 Cover the medical device with the KaVo Cleanpac bag and place it on the corresponding care adapter 4 Hold the c...

Page 22: ...adaptations unless otherwise specified refer to the Instructions for Use of the washer disinfector 7 3 2 Automated drying The drying procedure is usually part of the cleaning programme of the washer d...

Page 23: ...in accordance with German Robert Koch Institute requirements KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each ster...

Page 24: ...e pack 7 6 Sterilisation Sterilisation in a steam steriliser autoclave in accordance with EN 13060 EN ISO 17665 1 CAUTION Improper service and care Risk of injury 4 Service regularly with suitable age...

Page 25: ...60 7 Reprocessing steps in accordance with ISO 17664 7 7 Storage 25 28 7 7 Storage Reprocessed products must be stored appropriately protected from dust in a dry dark and cool space with a low germ le...

Page 26: ...04 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 Replacement filter 1 002 0271 Wrench for water filter 1 002 0321 O ring 8 3x0 68 0 200 6120 KaVo CLEANspray...

Page 27: ...not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or con...

Page 28: ...1 015 0904 bd 20221122 01 en...

Reviews: