Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640
7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
20 / 28
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
4
Take suitable personal protective measures.
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
4
Reprocess the medical device right after treatment.
4
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
4
Remove the dental bur from the medical device.
4
Remove all residual cement, composite or blood immediately.
4
Do not place in solutions or similar substances.
7.2 Manual Reprocessing
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
7.2.1 Manual external cleaning
Accessories required:
▪ Tap water
▪ Brush, e.g. medium-hard toothbrush
4
Brush off under flowing tap water.
7.2.2 Manual internal cleaning
This product is not designed for manual internal cleaning.
For effective reprocessing, the inside of the device must be subjected to auto-
mated cleaning in a cleaning and disinfection unit in accordance with ISO
15883-1.