background image

Instructions for use

Spare cartridge for EXPERTtorque mini -

1.007.9458, 1.007.9457

Always be on the safe side.

Summary of Contents for 1.007.9457

Page 1: ...Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...afety instructions 2 2 Product description 3 2 1 Technical data 3 3 Exchanging the spare turbine 4 4 First use 6 5 Function test 7 6 Tools and consumables 8 Instructions for use Spare cartridge for EX...

Page 4: ...ment for the proper purpose to protect him or herself the patient and third parties from hazards to avoid contamination by the product 1 2 Safety instructions CAUTION Injury or damage due to wear Irre...

Page 5: ...t turbine EXPERTtorque mini with and without wrench Mat no 1 007 9458 and Mat no 1 007 9457 consisting of 1 x replacement turbine assembled 1 x wrench optional 2 x O rings 1 x Spring washer larger out...

Page 6: ...wrench toward the cover and unscrew the cover in counterclockwise direction Take out the replacement turbine Use tweezers to push out and take out the replacement turbine from the front Remove the O...

Page 7: ...t position of the O rings Functional problem Make sure that the O rings are positioned correctly in the recesses Replace both O rings Spray O rings with KaVo Spray and insert them in the head housing...

Page 8: ...r Ensure that the cooling air is dry clean and uncontaminated in accordance with EN ISO 7494 2 Spray the KaVo recommended care product through the medical product and attach the medical product to the...

Page 9: ...art the medical device Perform a test run under axial and radial load Tighten the cover again applying a torque of 100 Ncm 6 Ncm Note If the unit runs poorly or heats up in untypical manner please ret...

Page 10: ...ter diameter shiny 1 003 5900 Spring washer smaller outer diameter coloured brown 1 003 7576 Material summary Mat No Instrument stand 2151 0 411 9501 Insert for medical device 0 411 9902 Cleanpac 0 41...

Page 11: ......

Page 12: ...1 008 6172 kb 20130114 02 en...