![KaVo 1.003.7712 Instructions For Use Manual Download Page 47](http://html.mh-extra.com/html/kavo/1-003-7712/1-003-7712_instructions-for-use-manual_658087047.webp)
▶
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
7.4 Drying
Manual drying
▶
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Machine drying
The drying procedure is normally part of the cleaning program of the ther‐
modisinfector.
▶
Follow the instructions for use of the thermodisinfector.
Preparation methods according to ISO 17664