Directly after automated cleaning, treat the medical device with the
care products and systems provided by KaVo.
6.2.3
Cleaning: Manual cleaning of the inside
Can only be done with KaVo CLEANspray or KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place
it on the corresponding care adapter. Press the spray button three
times for 2 seconds each time. Remove the medical device from the
spray attachment and let the cleaner work for one minute.
Preparation methods according to ISO 17664
51